NOTE:
To verify the electrical connector is secure, pull on the gray check connection area
disconnect.
1
2
3
Figure 4-5. Connection of Junction Box Harness Electrical Line
1
Alignment Mark
2
Red Alignment Dot
3
Check Connection Area
2. Position the opposite end of the Junction Box Harness next to the front of the Console.
a.
Secure the safety cable at the end of the gas line to the safety cable ring on the
Console.
b.
Remove the protective covers from the Junction Box Harness connectors and the sockets on
the Console.
c.
Connect the gas connection hub on the Junction Box Harness to the gas connection socket
on the Console (Figure 4‑3). To make the connection, align the ridge (callout 3) and red
unlocked position marker (callout 2) on the gas connection hub with the alignment mark
(callout 1) above the gas connection socket, push in the gas connection hub, and turn the gas
connection hub counterclockwise to the locked position (callout 4). A clicking sound indicates
a successful connection.
d.
Connect the fiber optic connector on the Junction Box Harness to the fiber optic socket on the
Console (Figure 4‑4). To make the connection, align the alignment mark (callout 1) above the
fiber optic socket with the red alignment dot (callout 2) on the fiber optic connector and push
in the connector. A clicking sound indicates a successful connection.
NOTE:
To verify the fiber optic connector is secure, pull on the check connection area identified
e.
Connect the electrical connector on the Junction Box Harness to the electrical socket on the
Console (Figure 4‑5). To make the connection, align the alignment mark (callout 1) on the
electrical socket with the red alignment dot (callout 2) on the electrical connector and push in
the connector. A clicking sound indicates a successful connection..
NOTE:
To verify the electrical connector is secure, pull on the gray check connection area
disconnect.
f.
Direct the Junction Box Harness towards the floor to minimize the potential for tripping.
Summary of Contents for Visual-ICE MRI
Page 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Page 20: ...1 10 User Manual This page intentionally left blank...
Page 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Page 56: ...4 14 User Manual This page intentionally left blank...
Page 66: ...5 10 User Manual This page intentionally left blank...
Page 80: ...9 2 User Manual This page intentionally left blank...
Page 118: ...12 2 User Manual This page intentionally left blank...
Page 120: ...13 2 User Manual This page intentionally left blank...
Page 121: ......