Observations

It is essential to always introduce clean and dry tubes.

Avoid drops in tubes’ locations, when placing one tube will wet the sensor and this will induce
errors in the system.
If by chance the system detected drops from distillation stucked in the glass wall of the tubes,
this will try to eliminate them. If after 3 attempts it is not achieved the system will show error by
making a long pip and turning front led into blinking red.To get operative mode again just press
the button once [0 state], cyan led, then clean and dry tubes again.
If operation would like to be cancelled, in distillation mode (2 and 3 state), press the button
during 2 seconds.
In [0 state] press the button during 2 seconds to stop the system (standby mode).To wake up
standby mode just press the button once.
Also valid for distillators of single element by alcohol heater (see photo).

Once 3.2ml of sample are distilled the system will return automatically to [0 state] turning off
the heater and showing it by a double pip and with front led once again turning iinto cyan. At
this point the content of the 3.2ml tube can be titrated by introducing its content in an
erlenmeyer flask.
For a new analysis

and place them again at their locations with the help

of the button repeating previous steps beginning from

.

clean and dry the tubes

step 08

Working method

12

13

02

01

03

04

05

07

07

08

09

09