GAB tenamatic Quick Start Manual Download Page 1

01
02
03

04

Characteristics

Instructions

Assembly and previous operation

GAB Sistemática Analítica S.L.

Sant Jordi, 30. Moja-Olèrdola 08734. Bcn [Spain]

Tel. +34 938 171 842

Fax. +34 938 171 436

www.gabsystem.com

[email protected]

Place the Volatimeter over the Tenamatic (see backwards photo).
Join a silicone tube from the outflow of the cooler to the inox tube of the inox arc.
Connect with the bridge power cable from the left backward plug of the Tenamatic to the power
plug of the Volatimeter.

Prepare the flask with 11ml of sample to analyze, 1-2 granules of siliconed pumice stone and
place it over the Volatimeter heater.
Open the water key to cool the Volatimeter.
Switch on the Tenamatic backward’s connector and make sure that the one from Volatimeter is
also switched on (1 position).The front of the Tenamatic will shine in cyan color [0 state]

Connect the power cable from the right backward plug from the Tenamatic to the network.

Change or place a clean and dry tube in the 3.2ml location.

[1 state]

[2 state]

ront led become purple

Press the front button once, the front light will shine in dark blue colour

. In this state

change or place a clean and dry tube in the 5.1ml location.
Press again the button and the front led will shine in red colour

.At this state the sample

previously prepared starts heating.
After few minutes the sample will boil and distillation will begin. When level of 5.1ml is reached
the system will turn automatically 90º clockwise and [state 3] will begin, f
and will center 3.2ml tube under the center of the arc.

Working method

08
09

10

11

05
06

07

· Compatible with electric Volatimeter GAB
· Automatic test tube swaping system
· It stops the heater at the end of the distillation without the user intervention
· Exact dosification through optical sensors: it guarantees result precision and its reliability
· Unattended operation
· Single button control
· It saves a notable amount of time and energy
· Drastically reduction of flasks breakage
· Power: 110-230 V / 50-60 Hz

Tenamatic GAB (ref. 1010016)

Equipment configuration

·

·

·

·

·

1 inox structure
4 tenamatic test tubes 15x100
1 tenamatic conection cable
1 small silicone tube + 2 joints
1 instructions manual

ref. 3026212
ref. 36.760-180-BP

· GAB volatimeter

ref. 1010001

Related products