![G-Tek LM-U PRO Series Operating Manual Download Page 8](http://html1.mh-extra.com/html/g-tek/lm-u-pro-series/lm-u-pro-series_operating-manual_3087402008.webp)
Page
7
Memory Type
Flash, Non-volatile, Data Retention up to 100 years
Download Batch
Intermediate after any time the batch started/After the
Batch is stopped.
Environmental
Temperature
(Operation)
5˚C to 45˚C
(Limiting) 0˚C to 50˚C
(Storage) -
20˚C to 60˚C
Humidity
(Operation) 10 to 80 % RH Non-Condensing
(Storage) 5 to 90 % RH Non-Condensing
Power Requirement
Supply Voltage (Battery
Operated)
3 V, 3500 mAh, 2 X AA (1.5V) battery; User Replaceable
Battery Life
Up to 1 year battery operating life with shelf life of 1 year (if
data is stored at 15 minutes interval and display OFF mode).
The battery indicator on the display provides information on
the remaining lifetime.
Battery Reverse Polarity
Protected
General
Restore Calibration
User can revert to original factory calibration any time
Reusability
Yes
Online Feature
Power saver mode when USB connected, Online Batch with
Current Data Display using LMView software
Communication
Connectivity
Micro USB data cable (USB 2.0, USB 3.0 Compatible)
Data Download Time
Approx. 2 min for 32000 data records
Standards
EMI-EMC
EN 61326 Class A
Pollution Degree
II
Installation Category
I
Vibration
2g Peak (10Hz-150Hz)
Shock
IEC 61010-1
IP Rating
IP 30
Overall Dimension
Dimension L x W x D
(mm)
135 (L) x 76 (W) x 27 (D)
Mounting
Wall Mounting Bracket
*Feature depends on the options ordered; Specifications are subjected to change without notice.
Table 2 Sensor Types with their respective ranges