4.
When using Pistonphone Type 42AP
(42AA), the correction factor for this setup
is -0.62 dB. Therefore, your analyzer
should ideally read:
114.00 dB - 0.62 dB = 113.38 dB
I/O button
6.3
Calibrating IEC 60711 Ear Simulator Coupler RA0045
Note:
Requires a Calibration Adapter RA0157 (must be ordered separately).
LED
Display
2.
Mount the Calibration Adapter
RA0157 into the pistonphone:
1.
Loosen the pistonphone’s retention
collar.
3. a. Switch on the pistonphone via the
I/O
button.
b. Check that the LED is showing
green.
c. Place the pistonphone (Type 42AP
shown) carefully onto the ear canal.
Note:
Hold the pistonphone strictly
horizontally.
Fig. 6.12
Fig. 6.13
Fig. 6.14
G.R.A.S. SOUND & VIBRATION
Page 27
Hearing-Protector Test Fixture Type 45CA