Insulation Resistance Test 

When measuring, press the 

Type Select

 button to select the test type as 

Insulation 

Resistance

,  and  clamp  the  measured  insulation  resistance  like  clamp  the  varistor  as 

described above.Press the 

Start Test

 button to start the test. In the lower right corner of 

the screen, there is a 

in progress...

 message. The test will stop automatically after about 

15 seconds from the start of the test. If you want to interrupt the test in advance, you can 
press the 

Exit

 button directly.

 

6

Backlight Control 

After  booting  up,  press  and  hold  the 

Exit

  button  to  enter  the  backlight  brightness 

adjustment page. Press the 

 and 

 arrow keys to adjust the backlight brightness. 

7

Qualification Judgment Setting 

After booting up, press the 

OK

 button to enter the qualification judgment mode. At 

this time, you can see that the bottom text box on the screen is flashing, indicating that the 

item can be set.Press the "

" or "

" (short press plus or minus 1 when the upper and 

lower limits are adjusted, long press plus or minus 50) to adjust the parameters, press the 

"

" or "

" key to change the cursor position.When the qualification judgment state is 

adjusted  to  the  "on"  state,  the  qualification  judgment  can  be  turned  on,  otherwise  the 
qualification judgment is turned off.The right side of "U:" is the judgment voltage range, 
the value in the left box is the lower limit of the judgment, and the right is the upper limit 
value.When  the  test  is  completed,  if  the  measured  voltage  is  within  this  range,  it  means 
qualified.

Pass

  will  be  displayed  in  the 

Status

  box  and  the  buzzer  will  sound 

once,otherwise, 

Fail

 is displayed and the buzzer sounds three times. As shown below:

 

 
 
 
 
 
 
 
 

Summary of Contents for ES9020

Page 1: ......

Page 2: ...cations 4 V Structure 7 VI Operation Method 7 1 Power On Off 7 2 Battery Voltage Check 7 3 Varistor Test 8 4 Discharge Tube Test 9 5 Insulation Resistance Test 10 6 Backlight Control 10 7 Qualificatio...

Page 3: ...nt is in use and the enclosure or test wire is broken and the metal is exposed please stop using it Do not place and store the instrument for a long period of time under conditions of high temperature...

Page 4: ...erformance and strong in anti interference ability Moreover it has anti vibration dust proof and moisture proof structure and adapts to the harsh working environment It is an indispensable instrument...

Page 5: ...nd leakage current measurement insulation resistance test Ambient temperature and humidity 23 5 below 75 rh Power 7 2V Large capacity rechargeable battery Measure method DC voltage drop method Shift A...

Page 6: ...ee times Battery voltage The low battery voltage symbol indicates that the battery is promptly replaced when the battery voltage is low Automatic shut down The splash screen indicates that it will aut...

Page 7: ...3000V rms between the circuit and the housing Electromagnetic properties IEC61326 EMC Suitable for safety regulations IEC61010 1 CAT 300V CAT IV 150V pollution level 2 V Structure 1 LCD 2 Function but...

Page 8: ...Type Select button to select varistor and clamp the varistor under test as shown below There are two modes to choose from under the varistor type namely automatic mode and manual mode which can be sel...

Page 9: ...under test If you want to end the test you can press the Exit button or press the Start Test button when checking the varistor leakage current value 4 Discharge Tube Test When measuring press the Type...

Page 10: ...ting up press the OK button to enter the qualification judgment mode At this time you can see that the bottom text box on the screen is flashing indicating that the item can be set Press the or short...