Body Fat Measuring Procedure in ‘Stand-Alone’ Mode 
 

1.

 

Turn Power Switch ON.  The display will read "FUTREX-6100/XL", followed by the 
software identification number, then "Advanced Body Composition Analyzer". 
 

2.

 

The FUTREX-6100/XL must be Zero Adjusted. 

a.

 

Insert Light Wand into Optical Standard. 

b.

 

Press the ZERO ADJUST key.  The display will flash "Reading...".  Once the 
zero adjustment has been completed, the display will read "READY". 

c.

 

Remove the Light Wand from the Optical Standard.  Then install the Light 
Shield on the Light Wand (allow the Light Wand to protrude from the edge of 
the shield by about 1/4 of an inch).  Line up the silver stripe on the Light 
Wand with the silver stripe on the Optical Standard. 
 

 

   

 

Figure 3 

 

3.

 

To start the measurement sequence, press ENTER. The instrument will then request an 
ID number – this is only needed for field testing the calibration of the devices and is used 
infrequently. Simply press ENTER again to continue.  

 

4.

 

Enter sex, date of birth (MM/DD/YY), height in inches, and weight. The display will then 
say “Ready to Read”  

NOTE:  It is important to line the silver colored 

stripe on the Optical Standard with the silver 

stripe on the Light Wand.  These alignment 

marks ensure that the Optical Standard is read 

exactly the same each time. 

NOTE:  The Light Shield is provided to prohibit 

external light from interfering with the 

estimations of percent body fat.  Therefore, this 

shield must always be used when making a 

measurement. 

Summary of Contents for FUTREX-6100

Page 1: ...1 FUTREX 6100 Owner s Manual Revision 6 0 Revision Date January 20 2020 Futrex Spotcheck Inc 130 WESTERN MARYLAND PARKWAY HAGERSTOWN MD 21740 1 301 733 9368 FAX 1 301 733 4660 E mail info futrex com...

Page 2: ...6100 Page 5 Service Requirements Page 5 Appearance and Part Names Page 6 Setting Up the FUTREX 6100 Page 7 Body Fat Measuring Procedure in Stand Alone Mode Pages 8 9 Locating the Biceps Measuring Site...

Page 3: ...d healthcare provider The FUTREX 6100 is not intended to be a diagnostic device and is for screening purposes only Read all of the information in the Owner s Manual before operating the device Operate...

Page 4: ...arranty Information The following warranty card is included with your FUTREX 6100 Please read the card and return it to Futrex Support is available by calling 301 733 9368 or E Mailing service Futrex...

Page 5: ...nty period or following please contact Futrex Support and Service at 301 733 9368 or via e mail Service Futrex com Futrex is the only authorized service representative for Futrex products DO NOT attem...

Page 6: ...rt Names The FUTREX 6100 consists of the following items 4 AC Adapter 5 Optical Standard 6 Light Shield 7 USB Cable 8 Biceps Locator 9 Fitness Wheel 2 Light Wand 3 FUTREX 6100 Software 1 Main Device C...

Page 7: ...y Be sure the batteries are oriented as indicated in the battery compartment 2 Connecting the Light Wand Ensure the Light Wand Connector is attached to the Main Body Connector NOTE Futrex recommends a...

Page 8: ...lver stripe on the Light Wand with the silver stripe on the Optical Standard Figure 3 3 To start the measurement sequence press ENTER The instrument will then request an ID number this is only needed...

Page 9: ...Reading 9 Percent body fat will then be printed and displayed 10 Press CLEAR to perform another measurement 11 When finished turn power switch to OFF CAUTION Do not move the Light Wand while the word...

Page 10: ...Inc Admin Functions on the Stand Alone FUTREX 6100 There are a few back end functions on the FUTREX 6100 that can be adjusted depending on the region the device is being used in or the mode it is to...

Page 11: ...date format stored within the FUTREX 6100 XL a Press the SELECT PROGRAMS Key b Press the DATE Key and step to the DATE selection Press ENTER and the current date will be displayed c To change displaye...

Page 12: ...o toggle between the valid decimal delimiters and thousands separators 9 999 99 and 9 999 99 respectively Figure 5 and Figure 6 Futrex Connect Sample Results Printouts IMPORTANT Please set up your Fut...

Page 13: ...H non condensing Operating Temperature Humidity Barometric Pressure Range Temperature 50 F to 104 F 10 C to 40 C Humidity 15 90 RH non condensing Barometric Pressure 105kPa to 80 kPa 790 600 mmHg Tabl...

Page 14: ...le Accurate Measurements document b Use an alcohol swab to gently clean the light wand c Take the measurement on the other bicep if possible d Ensure that the probe is fully seated in the optical stan...

Page 15: ...the product does cause harmful interference to radio or television reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one...

Page 16: ...16 Futrex Spotcheck Inc 130 WESTERN MARYLAND PARKWAY HAGERSTOWN MD 21740 1 301 733 9368 FAX 1 301 733 4660 E mail info futrex com...

Reviews: