background image

LCF Tension and Compression Sensor Family Manual

10

Sensor Solution Source

Load · Torque · Pressure · Multi-Axis · Calibration · Instruments · Software

www.futek.com

Shield Usage and Connections

•  Cable shielding should be grounded on one 

end, either the sensor side or instrument 
side to avoid ground loops .

•  A shield connection listed as floating on a 

sensors spec sheet means the cable shield is 
not connected on the sensor side and may 
be connected on the instrument side 
to ground .

Calibration

•  A yearly calibration is recommended .  But 

verification and calibration period shall be 
defined based on application, conditions, 
endurance and usage .

•  FUTEK offers NIST calibrations as well 

as A2LA accredited calibrations for total 
uncertainty .

•  For more information on available 

calibrations visit FUTEK calibration web 
page at: 

http://www .futek .com/calibration-

services .aspx

•  For recalibration orders visit the FUTEK 

recalibration page at: 

http://www .futek .com/

recalibration .aspx

•  An online summary of calibration results 

is available at: 

http://www .futek .com/

calibrationData .aspx

SHUNT

A shunt is an external resistance applied across 
two points on the load cell’s Wheatstone 
bridge to generate a known, fixed output from 
the sensor .

Shunt results can be used to set up instruments 
as well as compare changes to the load cell 
output over time and usage .

When selecting the appropriate shunt 
resistance for your load cell, we recommend a 
resistance that generates an output of about 
80% of the sensor’s rated output . It is important 
to have a shunt resistance that results in an 
output that is less than the full output of the 
load cell .

An online shunt calculator can be found at 

http://www .futek .com/shuntcalc .aspx

 to find 

a resistance that will generate a certain shunt 
output level, or to estimate the output for a 
known shunt resistance .

Additionally, recommended shunt resistance 
levels may be available on the sensor spec 
sheet .

TEDS

Transducer Electronic Data Sheet (TEDS) 
IEEE1451 .4 standard is available for FUTEK 
sensors and is utilized by select FUTEK 
instruments .  

Through the use of TEDS load cell calibration 
information can be stored with sensor, or 
sensor cable, for use with TEDS capable 
instruments .

FUTEK utilizes the Bridge Sensor template 33 
for the LCF family .

The following FUTEK instruments are TEDS and 
LCF compatible:

IPM Series

  

Panel Mount Display

IHH Series

  

Handheld Instrument

Bridge Sensor

XXX

+ Excitation

+ Signal

Sh

un

t C

al

– Signal

– Excitation

Shield

Power Supply

Shield

Jacket

Shield

Wires

Summary of Contents for LCF Series

Page 1: ...Sensor Solutions Source Load Torque Pressure Multi Axis Calibration Instruments Software www futek com LCF Series Tension and Compression Sensor Family Manual...

Page 2: ...ments Software www futek com Table of Contents Features 3 Mechanical Installation 4 Maximum Installation Torque 5 Mounting and Installation 6 Bolt Torque and Tightening Pattern 7 Cable Care and Routin...

Page 3: ...ection of capacities ranging from 25 lb to 100 000 lb Internal amplifiers for voltage or current output available on select models Integrated TEDS calibration information available on some models Fati...

Page 4: ...be used to assist in determining if the sensor can withstand any unavoidable off axis loads and moments Extraneous load information can be found at http www futek com extraneous load factor An extran...

Page 5: ...IMUM BOLT TORQUE in lbs LCF300 25 100 0 50 100 100 0 250 500 100 0 LCF400 250 500 700 0 1000 700 0 2500 700 0 5000 700 0 LCF450 LCF455 300 700 0 500 140 0 1000 450 0 2000 1000 0 5000 2100 0 10000 2100...

Page 6: ...ring Further mounting suggestions Mounting and Installation Below is information for proper mounting and installation Refer to the sensor spec sheet for thread information and proper load cell orienta...

Page 7: ...g or heat treating is performed after grinding on the mounting plate re verify that the surface maintained its flatness Heating of materials can cause warpage Make sure that the mounting plate can sup...

Page 8: ...us of ten times the diameter of the sensor cable for dynamic or moving applications and not exceed a onetime static permanent bend of two to three times the diameter of the cable CABLE JACKET REFERENC...

Page 9: ...ctions Additional connection standard colors are Orange and Blue Consult the sensor s online spec sheet for any further wiring information LCF EXCITATION POWER LEVELS SENSOR FAMILY MAX EXCITATION LCF3...

Page 10: ...atstone bridge to generate a known fixed output from the sensor Shunt results can be used to set up instruments as well as compare changes to the load cell output over time and usage When selecting th...

Page 11: ...oading surface Sensor loading surface is not obstruct ed or supported and able to flex under load Sensor support is not giving while sensor is loaded Internal disconnections or shorts would not be ava...

Page 12: ...le or noisy Sensor is exposed to temperature change Sensor support is not rigid and firm Sensor exposed to liquid or humidity Friction in assembly Stability of power supply and noise levels Fixtures f...

Page 13: ...and calibration period shall be defined based on application conditions endurance and usage Calibration data may be available online at http www futek com calibrationData aspx To send in your sensor o...

Reviews: