8-6
Chapter 8 Brachial Blood Pressure Examinations
when the first measurement is completed. The second
and third measurements start automatically when the
specified time has elapsed after the start time of the
previous measurement. To stop a measurement, press the
[STOP] key. The blood pressure measurement is aborted
immediately.
6
When the measurement is completed,
the examination result will be displayed.
• The results of the last three blood pressure
measurements are displayed.
• The average of the last three blood pressure
measurements is displayed at the bottom of the display.
• If there are any blood pressure measurement results
that should not be used to calculate the average blood
pressure, remove the check from the check box. The
average blood pressure is calculated from the checked
results.
Examination Result Display
When the examination is completed, touch [Use] to display the examination result.
• The brachial blood pressure measurement(s) and comments are shown at the top of the display.
• Average systolic blood pressures, diastolic blood pressures and pulse rates for the right and left brachia are shown at the
center of the display.
• The last three right and left systolic blood pressures, diastolic blood pressures and pulse rates for this ID are shown at the
bottom of the display.
●
Printing Conditions of Observation
• High blood pressure.
The systolic blood pressure for the left/right brachium is
above the high blood pressure threshold (SYS 2)
The diastolic blood pressure for the left/right brachium
is above the high blood pressure threshold (DIA 2)
• Difference between both side of BP.
The difference between the left and right systolic blood
pressures is above the left and right blood pressure
difference threshold
• The blood pressure is at the high end of the normal
range.
The systolic blood pressure for the left/right brachium is
above the high blood pressure threshold (SYS 1)
The diastolic blood pressure for the left/right brachium
is above the high blood pressure threshold (DIA 1)
●
[Save]
Touch this to display the "Storage Media" window. Select a
media to save the data.
Summary of Contents for VaSera VS-2000 Series
Page 4: ......
Page 22: ...18 ...
Page 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Page 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Page 172: ...9 22 Chapter 9 File Transfers Communication History ...
Page 176: ...10 4 Chapter 10 Recording Daily Reports ...
Page 214: ...12 12 Chapter 12 Maintenance and Inspection ...