Chapter 14 Troubleshooting
Troubleshooting
Troubleshooting
This section explains the troubleshooting for each case.
Other than the troubles stated below, troubles of the bedside monitor, telemetry transmitter, or other device can be
considered.
Refer also to the operation manual of those devices.
Wired Network (DS-LANIII), TCP/IP Network
Telemetry, TCON
Situation
Cause
Solution
The waveforms and numeric data
for the wired network beds are not
displayed.
The central ID is duplicated.
Make sure to set a unique central ID for each central
monitor. (1 to 16)
(
"Central ID" P7-28)
The Bed ID of the bedside monitor
or LAN-ID of the telemetry receiver
is duplicated.
Make sure to set a unique Bed ID for the bedside monitor
and LAN-ID for the telemetry receiver.
The administrator does not exist.
One of the central monitors must have the Central ID, "001"
in a network system. (
"Central ID" P7-28)
A HUB not compatible to wired
network is used.
For the DS-LANIII network, use the recommended
switching HUB.
Do not confuse the HUB for DS-LAN and TCP/IP network.
A central monitor which is not
compatible is used.
The following central monitors can not be used with the
DS-LANIII network.
DS-5700
DS-5800N/NX/NXMB
DS-7600/7600W with software version V05 and prior
DS-LAN Cable (Ethernet branch
cable or connection cable) is not
correctly connected. Or, the wire is
broken.
Check if DS-LAN cable is properly connected. Or, replace
the DS-LAN cable.
The network connecting device is
malfunctioning.
Replace the device.
A monitor with DS-LANII setting is
used.
Make sure to set DS-LANIII for all the monitors.
Situation
Cause
Solution
The waveform transmission is often
interrupted.
A low battery mark is displayed in the
waveform area for the telemetry receiving
bed.
Replace the transmitter battery with a new one.
The patient is located too far from the
receiver antenna.
Check the antenna system.
There is a metallic obstruction (elevator,
door, etc.) between the transmitter and
receiver.
Try to prevent metallic obstruction between the
transmitter and receiver.
A noise is interfering on the
waveform, and the waveform
suddenly changes.
[OFF] is selected for the AC filter.
Select [ON] for "AC Filter" under ECG setup.
(
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......