7-18
Chapter 7 Alarm Function
ECG Alarm at Lead-Off Condition
ECG Alarm at Lead-Off Condition
When ECG lead is detached, some waveforms may become immeasurable depending on the detached lead.
In such case, ECG waveform or respiration waveform will be displayed as baseline, and ECG related alarm will
generate.
ECG related alarms during Lead-Off condition are as follows.
HR Alarm
Arrhythmia Alarm
ST Alarm
RR Alarm of Impedance Respiration
Apnea Alarm of Impedance Respiration
If the alarm generated during lead-off condition is considered not reliable, selecting [OFF] for "Alarm Judgment"
(Initial Settings > Alarm Setup > During Lead OFF) will not generate the ECG related alarm during lead-off
condition.
For the alarm function during lead-off condition, the following setup can be performed on the "Alarm Setup".
ON/OFF of Alarm Judgment
ON/OFF of Lead OFF Message
ON/OFF of Alarm Printing
Lead OFF Alarm Interval (5/30/60 sec.)
WARNING
If the "Alarm Judgment" for "During Lead OFF" is set to OFF, HR alarm and arrhythmia alarm
will not be generated at lead-off condition. If this condition is left unresolved, a sudden
change of the patient may not be noticed. Take prompt action when the lead-off condition is
detected.
The ON/OFF setting of "Alarm Judgment" for "During Lead OFF" is effective for telemetry
beds only. For the wired network beds, HR alarm and arrhythmia alarm will not generate
regardless of this setting.
NOTE
These setups should be performed by our service representative or system administrator of
your institution.
(
Maintenance Manual "Alarm" P7-2)
The settings will be synchronized with the administrating central monitor.
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......