7-2
Chapter 7 Alarm Function
General Description
Classification and Level of the Alarm
The classification and level of the alarms are explained below.
Alarm Classification
The alarms displayed on this equipment are classified as follows.
Alarm Level
The alarms are classified to Level S (top priority), Level H (high priority, urgent), Level M (medium priority,
caution), Level L (low priority, status), and Notification, and the message will be displayed according to the priority
of Level S > Level H > Level M > Level L > Notification.
The displayed messages will flash in red and white for Level S, red for Level H, yellow for Level M, blue for Level
L, and white for Notification.
CAUTION
When more than one alarms of the same priority are generated, the newer alarm message
will be displayed.
Alarm Classification
Details
Vital Alarm
Alarms generated based on vital information. Includes numeric data alarms and arrhythmia alarms.
Arrhythmia Alarm
Alarms based on arrhythmia analysis. (Ex.: Asystole, VF)
Numeric Data Alarm
Alarms based on numeric data. (ex.: Upper HR, Lower BP1)
Equipment Alarm
Alarms based on measurement status.
Includes alarms of arrhythmia status, measurement status, external equipment.
Arrhythmia Status
Arrhythmia analysis condition such as noise, etc.
(Ex.: ECG Low, ECG Artifact)
Measurement Status
Measurement status condition such as noise, etc. (Ex.: Lead-OFF, Check SpO
2
Sensor)
External Equipment
Alarm
Alarms output from external equipment. (Ex.: Ventilator Alarm)
System Status
Alarms related to the main unit and connected equipment. (Ex.: Check PHS comm., Check Data
Server Comm.)
Printer Status
Alarms related to printer. (Ex.: Check Cassette)
Alarm Priority, Level
Details
Alarm Sound
Displayed Color
Top Priority
S
Top Priority Alarm
Continuous
Red/white
High Priority
H
Life Threatening Alarm Continuous
Red
Medium Priority
M
Cautionary Alarm
5 seconds interval
Yellow
Low Priority
L
Treatment Needed
Alarm
15 seconds interval
Blue
Notification
N
Notification Alarm
Display Only
White
Summary of Contents for DYASCOPE 8000 Series
Page 1: ......
Page 2: ......
Page 12: ...Contents ...
Page 20: ...Contents ...
Page 44: ...1 2 Chapter 1 General Description Contents ...
Page 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Page 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Page 56: ...3 2 Chapter 3 Description of the Display Contents ...
Page 70: ...4 2 Chapter 4 Basic Operation Contents ...
Page 82: ...5 2 Chapter 5 Preparation Contents ...
Page 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Page 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Page 110: ...7 2 Chapter 7 Alarm Function Contents ...
Page 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Page 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Page 166: ...9 2 Chapter 9 Data Review Contents ...
Page 190: ...10 2 Chapter 10 Waveform Review Contents ...
Page 206: ...11 2 Chapter 11 Calculation Contents ...
Page 212: ...12 2 Chapter 12 Printing Contents ...
Page 230: ...13 2 Chapter 13 Menu Items Contents ...
Page 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Page 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Page 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Page 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Page 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Page 322: ...16 2 Chapter 16 Accessories Contents ...
Page 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Page 328: ...17 2 Chapter 17 Specification Contents ...
Page 342: ...17 14 Chapter 17 Specification Technical Information ...
Page 347: ......
Page 348: ......
Page 349: ......