Fukuda Denshi HBX-800 BIS Series Instruction Manual Download Page 2

 

 

Maintenance 

- 2 - 

Cleaning the Housing

 

Clean the housing using tightly squeezed gauze or an absorbent cotton 
cloth dampened with alcohol or a neutral cleanser. 
 

CAUTION 

z

  Clean the equipment frequently so stains can be removed easily. 

z

  To prevent injury, it is recommended to wear gloves when 

cleaning the equipment. 

z

  When sterilizing the entire room using a spray solution, pay close 

attention not to have liquids get into the equipment or connectors.

z

  Do not use organic solvents, thinner, toluene or benzene to avoid 

damaging the resin case. 

z

  Do not polish the housing with abrasive, chemical cleaner, 

alkaline or acidic detergent. Otherwise, the surface resin or paint 
coating may be damaged, resulting in discoloration, scratches, 
and other problems. 

 

Cleaning and Disinfecting the BISx Module 

z

 Clean any spillage of blood or solutions on BISx as soon as possible. 

z

 Use lint-free absorbent towels for spill cleanups. 

z

 Dampen the towel with detergent and lukewarm water to aid in 

cleaning. 

z

 After cleaning, wipe the PIC connector ends with alcohol and allow to 

dry completely 

z

 Residual moisture inside the connector may affect BISx performance. 

z

 Use lint-free absorbent towels dampened with a 10% bleach solution, 

or a commercial disinfectant.   

 

WARNING 

z

  Clean the equipment frequently so stains can be removed easily. 

z

  Universal Precautions shall be observed to prevent contact with 

blood or other potentially infectious materials.   

z

  Place contaminated materials in regulated waste container. 

z

  Whenever an event such as spillage of blood or solutions occurs, 

re-test leakage current before further use. 

z

  Do not mix disinfecting solutions (e.g., bleach and ammonia) as 

hazardous gases may result. 

 

CAUTION 

Do not autoclave the BISx Module or HBX-800. Autoclaving will 
seriously damage both components. 
Avoid liquid ingress to the patient interface cable. Contact of fluids 
with the PIC sensor connectors can interfere with PIC performance. 

 

Daily Check 

Conduct the following daily check. 

z

 Check if the cables are properly connected. 

 

Periodic Check 

The safety check conformed to IEC 60601-1 must be performed at least 
once a year for this equipment. 
Periodic check of medical electronic equipment is mandatory to prevent 
failures and accidents and to ensure safety and reliability. 
Periodic inspection may be performed by the medical institution or by a 
third party by concluding a “Maintenance Contract”. For more details, 
contact your nearest service representative. 
 

NOTE 

If any damage is found, contact your nearest service representative. 

 

Troubleshooting 

For troubleshooting, refer to the operation manual of the DS-8500 
System or other equipment which the HBX-800 is connected to. 
 

 

Accessories 

Accessories 

In order to satisfy product performance requirements, always use the 
accessories specified by Fukuda Denshi. 
 

No. Item 

Q’ty 

Operation Manual (this manual) 

2 Clip 

CAUTION 

z

  Handling the clip 

Do not attach the clip to a pole or board arranged vertically. 
The unit may fall off when excessive force is applied. 

     

         

       

 

 

Optional Accessories 

The following products are available as optional accessories. Purchase 
them as required. For more details, refer to our product catalogue or 
contact your nearest service representative. 
 

No.

Item Parts 

Number 

1 BISx 

185-0145-AMS 

2 Patient 

Interface 

Cable 

 

186-0107 

BIS Extended Use Sensor 

186-0160 

BIS Pediatric Sensor 

186-0200 

BIS Quatro Sensor 

186-0106 

6 CJO-15RR0.65 

AUX 

Connection 

Cable  551024870 

7 CJO-15RR1.5 

AUX 

Connection 

Cable  551024880 

CJO-15RR3 AUX Connection Cable 

551024890 

 

Specification 

The following is the specification for this equipment. 
For the performance of each parameter, refer to the operation manual of 
the DS-8500 System or other equipment which the HBX-800 is 
connected to. 

 
Size/Weight 

Size 

:  36(W) × 87(D) × 91(H) mm

±

10mm 

Weight : 

0.20kg

±

0.05kg 

*not including the connector part and protrusion. 

 

Environmental Conditions 

Operation Temperature 

:  10 to 40

°

Operation Humidity 

:  30 to 85

 (non-condensing) 

Transport/Storage Temperature :  –10 to 60

°

Transport/Storage Humidity 

:  10 to 95% 

(within absolute humidity of 
0

/90%RH) (non-condensing) 

 

Safety 

Type of protection against electric shock: Class I Equipment   

(Depends on the connected equipment) 

Degree of protection against electric shock: 

Type BF Applied Part (BIS Measurement) 

 

Power Supply 

Voltage  :  DC+12V (Supplied from the connected equipment) 
 

 