Chapter 4 Connection to the External Devices
BIS Data Input
BIS Data Input
By connecting the A-2000/A-3000 BIS monitor (Covidien), the patient's recovery from anesthesia can be verified
by BIS (Bispectral Index) value.
CAUTION
Refer to the BIS monitor operation manual and set the SQI value above 15.
ASCII should be set to communicate with this system. Make sure that ASCII is set on the
BIS monitor communication setting. Refer to the BIS monitor operation manual for
procedures.
Securely connect the cable to the serial or status connector of the AC Unit (DSA-81) and the
connector of the BIS monitor.
Connecting the A-2000/A-3000 (Covidien)
CAUTION
When connecting this equipment and the BIS monitor, use only the specified connection
cable.
Make sure that the BIS monitor is connected to the specified connector on this
equipment.When connecting the cable, make sure that the power of this equipment and the
BIS monitor is turned OFF.
1
Connect the serial connector or Status II connector on the AC Unit (DSA-81) to the serial connector on the BIS
monitor using the BIS connection cable (CJ0-03RS4).
External Device Setup
To display the BIS monitor data, external device setup is required.
1
Press the [Menu], [Initial Settings], [External Device],
[AC Unit] keys.
The "AC Unit" setup menu will be displayed.
2
Select the port to connect the BIS monitor.
3
Press the [Other] key.
4
Press the [BIS] key.
BIS Monitor
Connection Cable
For Status II Connector
For Serial Connector
A-2000
CJ-407RI-70BIS
CJO-03RS4
A-3000
3
4
2
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......