Sight Inflation
When “Sight Inflation” is set to ON, the maximum blood pressure level will be estimated during the
inflation process, then 35mmHg will be added and the measurement will start. The inflation speed is
slow but allows detecting any sudden increase of blood pressure to prevent re-inflation and discomfort
for the patient.
When “Sight Inflation” is set to OFF, normal cuff inflation, which inflates to the target level set
according to the previous measurement result, will be performed.
N O T E
z
The Sight Inflation setting is valid only during the NIBP automatic
measurement.
z
The “Sight Inflation” will not function if the patient classification is
“Neonate”.
z
During continuous measurement / 1 minute measurement, normal cuff
inflation will be performed regardless of the Sight Inflation setting.
z
When performing manual measurement / measurement at alarm
occurrence, it will inflate to the fixed pressure (Adult: 180mmHg, Child:
140mmHg, Neonate: 110mmHg) regardless of the Sight Inflation setting.
1
Select ON/OFF for “Sight Inflation”.
ON will set the Sight Inflation ON.
OFF will set the Sight Inflation OFF.
●
Oscillation Graph Display
The oscillation graph can be set to be displayed inside the NIBP numeric data box.
1
Select ON/OFF for “Osc. Graph Display”.
ON will display the oscillation graph inside the NIBP
numeric data box.
OFF will not display the oscillation graph.
Oscillation Graph
6
Non
-In
va
sive Blood Pre
ssu
re
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...