Continuous Measurement
The NIBP measurement can be continuously performed for 12 minutes. When the continuous
measurement is completed, the measurement interval will be set to 2.5 minutes.
If any abnormality on the cuff hose, etc. is found during the continuous measurement, the
measurement will cease.
During continuous measurement, a provisional SYS (systolic pressure) will be displayed during the
measurement. It will not be displayed during the Quick Measurement (measurement duration of 20 to
25 seconds).
1
Press the
(NIBP Continuous Measurement) key.
2
Start the NIBP continuous measurement.
To cease the measurement, press the
(NIBP Start/Stop) key, or press again the
(NIBP Continuous Measurement) key.
“Continuous Measurement”
Provisional SYS