Non-Invasive Blood Pressure
This menu allows setting the NIBP monitoring condition.
Auto Mode
: Sets the automatic interval measurement and starts the 1-minute interval
measurement and continuous measurement.
NIBP Alarm
: Sets the ON/OFF of NIBP alarm and upper / lower limit of systolic, diastolic,
and mean BP.
NIBP Measurement at
Alarm Occurrence
: Sets the condition to perform NIBP measurement at alarm generation.
Setup
: Sets the NIBP monitoring condition.
C A U T I O N
z
Pay attention when measuring the NIBP of patient with bleeding disorders
or hypercoagulation. The cuff inflation constricting the arm may cause
petechia or circulatory failure with blood clot.
z
Check the patient’s condition constantly while measuring over a long
period of time with interval of 2.5 minutes or less. Also, periodically check
the blood circulation while performing periodic measurement over a long
period of time. Congestion or rash may occur at the measuring site.
z
For the following situation, measurements will be terminated.
y
When the measurement time has exceeded 160 seconds for adult and
child, 80 seconds for neonate.
y
When the inflation value has exceeded 300mmHg for adult, 210mmHg
for child, and 150mmHg for neonate.
z
If used with the incorrect patient classification, it will not only cause
erroneous measurement, but the inflating level for the adult may be applied
to child or neonate causing a dangerous situation to the patient.
6
Non
-In
va
sive Blood Pre
ssu
re
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...