4
-
67
Periodic Recording
The recording will be automatically performed with the selected interval. Periodic recording can be
performed on the central monitor connected on the wired network system.
N O T E
z
If the periodic recording was interrupted due to paper out, etc., only the latest data
will be printed when it becomes available again.
z
QRS judgment will not be printed for periodic recording.
1
Press the Menu
→
Settings
→
Record
→
Periodic Record keys.
The periodic recording setup menu will be displayed.
2
Select the output recorder.
Built-in will record on the DS-7000 built-in recorder.
Central will record on the central monitor recorder .
8ch will record on the 8ch recorder.
The central monitor recorder of the smallest central ID will
be used. 8ch recorder selection can be performed on the
recorder setup menu.
3
Select the waveform for recording.
Up to 3 waveforms can be selected. If the output recorder is
8ch recorder, up to 8 waveforms can be selected.
The waveform position will be automatically adjusted when
recording.
4
Select the periodic interval.
Interval recording or timer recording can be selected for
periodic recording.
Press the Timer key.
The recording will automatically start at the programmed time.
Select the time to start recording.
Press the Interval key.
The recording will automatically start with the selected interval.
If 5 min. is selected, recording will start at 10:00, 10:05, …
10:25.
If 60 min. is selected, recording will start at 10:00, 11:00, …
12:00.
4
Re
cording Se
tup
Summary of Contents for DynaScope 7000 Series
Page 38: ...xxxvi Blank Page ...
Page 126: ...4 78 Blank Page ...
Page 148: ...5 22 Blank Page ...
Page 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Page 322: ...7 78 Blank Page ...
Page 428: ...10 32 Blank Page ...
Page 466: ...11 38 Blank Page ...
Page 474: ...12 8 Blank Page ...