The <SpO
2
Pulse Search> message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution : Check if the light emitting and receiving parts of the sensor LED are aligned.
The <SpO
2
No Pulse Detect> message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not positioned
correctly.
Solution : Check if the light emitting and receiving parts of the sensor LED are aligned.
The <SpO
2
Communication Error> message is displayed.
Cause
: Communication error with the SpO
2
module has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed, a failure of the
equipment can be considered. Cease the measurement and contact our service
representative.
The <SpO
2
Motion Artifact> message is displayed.
Cause
: There is excessive body motion from the patient which disables correct
measurement.
Solution 1 : Have the patient lie still.
Solution 2 : Relocate the sensor, or change the sensor to which the body motion will have less
influence.
Temperature
The temperature measurement accuracy is out of range.
Cause
: The connector of temperature sensor is not clean.
Solution : Clean the connector.
The measurement data on the monitor is displayed as “
ー
ー
ー
.
ー
”.
Cause
: The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly inserted/attached.
The <TEMP Disconnected> message is displayed on the screen
Cause
: When the temperature sensor is disconnected during TEMP monitoring, this
message will be displayed.
Solution : Check if the temperature probe is properly inserted/attached.
The <TEMP Unknown Sensor> message is displayed on the screen.
Cause 1 : The 700 series temperature probe may be used.
Solution : Use the 400 series temperature probe for measurement.
Cause 2 : There is a contact failure of the temperature probe.
Solution : Check if the temperature probe is properly inserted.
The <TEMP Communication Error> message is displayed on the screen.
Cause :
Communication
error
with
the TEMP board has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed, a failure of the
equipment can be considered. Cease the measurement and contact Fukuda Denshi.
The <TEMP Out of Range> message is displayed on the screen.
Cause
: The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly inserted.
8
-
9
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...