Chapter 12 Daily and Periodic Checks
12-5
12-3-1 Safety Check (6 items)
y
In order to ensure safe handling of medical equipment and measuring instruments and to enable tests to be
performed safely, leakage current and other factors that have adverse effect on human body are prescribed
by the safety test standards.
When performing the check, assessment should be made by measuring the leakage current with a leakage
current measure that satisfies the measurement conditions for medical electrical equipment.
WARNING
If, during a safety inspection, a value that exceeds the allowable limit is confirmed,
immediately stop using the given unit; then replace the defective part or make
proper repairs. Failure to repair the unit can lead to a serious accident.
Note:
z
There are a several types of measurement states in the measurement of leakage current, be
sure to determine the maximum value among all the measured values and adopt it as the
measured value. It is also important to compare the measured values against those in the
past and check whether there are any problematic changes in them.
Summary of Contents for CardiMax FX-8322
Page 4: ......
Page 16: ......
Page 17: ...Chapter 1 Overview C Ch ha ap pt te er r 1 1 Overview 1 Overview 1 1 ...
Page 18: ...Chapter 1 Overview ...
Page 20: ...Chapter 1 Overview 1 2 ...
Page 22: ...Chapter 2 Specifications ...
Page 28: ...Chapter 2 Specifications 2 6 ...
Page 30: ...Chapter 3 Names and Functions of Parts ...
Page 34: ...Chapter 4 Operational Description ...
Page 48: ...Chapter 5 Internal Connection Diagram ...
Page 50: ...Chapter 5 Internal Connection Diagram 5 2 ...
Page 51: ...Chapter 6 Spare Parts List C Ch ha ap pt te er r 6 6 Spare Parts List 6 Spare Parts List 6 1 ...
Page 52: ...Chapter 6 Spare Parts List ...
Page 56: ...Chapter 6 Spare Parts List 6 4 ...
Page 58: ...Chapter 7 Disassembly and Assembly Procedure ...
Page 75: ...Chapter 7 Disassembly and Assembly Procedure 7 17 7 17 Outside Drawing ...
Page 76: ...Chapter 7 Disassembly and Assembly Procedure 7 18 ...
Page 78: ...Chapter 8 Software Update Procedure ...
Page 81: ...Chapter 9 Caution Labels C Ch ha ap pt te er r 9 9 Caution Labels 9 Caution Labels 9 1 ...
Page 82: ...Chapter 9 Caution Labels ...
Page 84: ...Chapter 9 Caution Labels 9 2 ...
Page 86: ...Chapter 10 Troubleshooting ...
Page 98: ...Chapter 11 Maintenance ...
Page 114: ...Chapter 12 Daily and Periodic Checks ...
Page 140: ......