QRG-04
Quick Reference Guide
Objective Lens
Cotton swab
Fiber
Cleaning before Splice Operation
Clamp chips
・
After installation, execute [Stabilized Electrodes] followed by an [Arc Calibration].
V-grooves
・
When the lens is dirty, clean it with a thin
cotton swab moistened with alcohol.
・
Remove the electrodes when cleaning the
lenses.
Cleaning Cautions
Electrode Replacement
・
When “Replace electrodes” message appears, or when the tip is damaged, replace electrodes.
・
Execute [Replace Electrodes]
・
Exchange to new electrode.
Install new electrodes
Loosen screws
Remove the old electrodes
Tighten screws
Screws
V-groove Clamps
Tools
Optical fiber cleaver
・
Clean rubber pads.
・
Clean rubber anvil.
・
Clean cleaving blade.
Ribbon fiber stripper
・
Clean stripping blade
・
Do not allow contact onto the electrode tips.
・
Use only 99% or better purity of alcohol
・
Use new lint-free cotton swab every time you clean.
Clean bottom of V-groove with a thin
cotton swab with alcohol.
Remove excess alcohol from V-grooves
with a clean dry swab.
Use a cleaved fiber end-face to dislodge.
《Replace the electrode after selecting [Electrode Replacement] in the maintenance menu.》