Problem
Solution
Lack of sensitivity
l
Ensure the anatomy being studied is in the focal
zone for the transducer.
Lack of penetration or
sensitivity
l
Ensure that there is adequate coupling medium (for
example, ultrasound gel) between the transducer
and the patient.
l
Adjust the position of the TGC sliders.
l
Ensure the appropriate transducer is being used.
Problem
Solution
One or more DICOM
services are not
working
l
If in the Network Status page, the location is set to
Not Connected
, select a location and restart the
system. See
l
Ask your IT Department if the IP address is correct
and if it is in range of the local network.
l
Ensure the services you need are selected with
check marks in the
DICOM Setup
page and that the
SC
(Storage Commit) and
MPPS
columns are also
checked for the selected archiver.
DICOM service
verification fails
l
Tap
Config
, and select a service that you would like
to check.
l
Select from the list and tap
Ping
.
l
If it has failed, the IP address is incorrect or the
destination server is not seen on the network.
DICOM troubleshooting
Troubleshooting
413
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...