9. Focal zone indicator
10. Status bar
Color box overlay
In Color Doppler Mode, the region of interest is located within the green color box
overlay.
Data is displayed in blue to display vascular flow moving away from the transducer
and red to display vascular flow moving toward the transducer.
The system applies the colors only to the image data within the color box.
Display maps and velocity scale
Two display maps are shown to the left of the image while in Color Doppler Mode. The
left column is the 2D display map. The right column is the color display map. It follows
the acronym BART color principle for Doppler imaging (Blue=Away from the
transducer, Red=Toward the transducer). In this principle, positive vascular flows are
indicated by colors in the red range. Negative flows are indicated by colors in the blue
range. Flow velocities for each direction increase from dark to light.
In Color Doppler, a velocity scale corresponding to the display map is shown to the
left of the display maps.
Priority indicator
Adjust the priority relationship between the overlay data and the background 2D data
so you can eliminate false readings, by using the
Priority
control.
Color Doppler Mode clinical display
251
Summary of Contents for VisualSonics Vevo MD
Page 1: ......
Page 2: ......
Page 12: ...12 ...
Page 69: ...System settings 69 ...
Page 70: ...70 System settings ...
Page 77: ...3 Tap DICOM Setup Connectivity 77 ...
Page 146: ...2 Tap User Management in the list on the left 146 System settings ...
Page 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Page 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Page 462: ...462 Index ...
Page 463: ...51370 01 1 0 51370 01 ...