4-12
Imaging
Controls in Color
To specify which controls appear in the Controls bar, see
appear only when the image is frozen.
Table 4-2: Color controls
Button name
Description
Available in
Live
Froze
n
Color
Allows you to select
Color
, Color Power Doppler (
CPD
), or
Variance
. On the P21xp transducer in the Cardiac exam type,
Variance
is available instead of
CPD
.
CPD
shows amplitude and is
useful for very low flow states.
Variance
displays a color map that
shows regurgitant jet turbulence. The highest velocities appear as
green.
Flow State
Lets you adjust the sensitivity of color parameters or settings to
enhance the flow state. The system initially selects a setting based
on the exam type.
High
optimizes for high flow states such as cardiac. The High
setting minimizes flash artifacts.
Medium
optimizes for medium flow states such as abdominal or
arterial flow.
Low
optimizes for low flow states such as breast or
musculoskeletal.
For a more precise setting, adjust the
Scale
control.
Scale
Sets the pulse repetition frequency (PRF). Tap this control, and
then tap
Up
or
Down
.
Wall Filter
Helps to filter out echoes from low frequency signals.
Tap to select the strength of the filter:
Low
,
Med
, or
High
.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...