2
S
TEP
10
Rinse
the shaft for at least
1
minute
in a large volume of sterile,
deionized water (for example, 8 L). Repeat
3 separate times
.
Discard
the rinse water
after each rinse.
S
TEP
11
Dry
the transducer with a
sterile, non-linting towel or medical-grade
air.
Examine
the transducer and cable for
any cracks.
DO NOT
use the transducer if
any damage is found.
Disinfecting
S
TEP
9
A
Place up to 90 cm of the shaft in an
approved
disinfectant. Make sure the controller cannot
fall into the solution.
CHECK
the expiration
date of the disinfectant, and follow
ALL
chemical manufacturer’s instructions for soak
time and temperature.
B Wipe
the controller and cable with a sterile,
non-linting cloth or single-use endoscopic
sponge moistened with disinfectant.
Remove
cleaner by wiping again with a
fresh sterile cloth or sponge moistened with
sterile water.
S
TEP
7
Set up the
equipment.
A
Insert the transducer into the
transducer adapter.
B
Insert the fork adapter
securely in between the
control wheels on the
controller and do not allow it
to swivel.
C
Connect the splitter to the
transducer adapter, the fork adapter, and the leakage tester.
D
Connect the conductivity probe to the leakage tester and insert
into 0.9% saline solution or Cidezyme/Enzol.
A
B
C
D
Testing for electrical leakage
S
TEP
8
Make sure the tester displays
Ready for Testing
, press
FULL
TEST
. Repeat the test two more
times.
If any of the results are
Fail
, do not
use the transducer and contact
FUJIFILM SonoSite for repair.
S
TEP
12
Transport and store
the
transducer in accordance with your
organization’s policies.
Transporting and storing
For complete instructions on cleaning, disinfecting,
transporting, and storing the TEE transducer, refer to
the
TEE Transducer User Guide
.
For complete instructions on using the leakage tester,
refer to the
ULT2000 Series User Manual
.
SonoSite is a registered trademark of FUJIFILM Sonosite, Inc. in various jurisdictions.
FUJIFILM is a registered trademark of FUJIFILM Corporation.
Value From Innovation is a
trademark of FUJIFILM Holdings America Corporation. All other trademarks are the property of their respective owners.
Copyright
©
2019 FUJIFILM SonoSite, Inc. All rights reserved.
P05497-09
EC Authorized
Representative
FUJIFILM SonoSite B.V.
Joop Geesinkweg 140
1114 AB Amsterdam,
The Netherlands
Manufacturer:
FUJIFILM SonoSite, Inc.
21919 30th Drive SE
Bothell, WA 9802, USA
T: 1-888-482-9449 or
1-425-951-1200
F: 1-425-951-1201
Australia Sponsor
FUJIFILM SonoSite
Australasia Pty Ltd
114 Old Pittwater Road
Brookvale, NSW, 2100
Australia