897N203029A FDC NX600 Instruction Manual 2022.01
3 Measurement
3.7 CRP Tests
NOTE
Each time a new box of CRP slides is used, the new QC card must be read.
NOTE
Calibration is necessary for CRP measurements under the following conditions.
• When the slide lot has changed
• When the validity term of calibration has expired (For the validity term, refer to the “In-
structions for Use” of the slides)
NOTE
If the validity term has expired, a “#” warning mark is printed along with the measurement
results.
NOTE
Specified diluent and mixing cups are necessary for CRP measurements.
NOTE
When the results of a measurement will exceed the measurement range, dilution can be
performed automatically at 21x or more by configuring the dilution factor at the sample in
-
formation registration screen. (When 2x has been set at the sample information registration
screen, the dilution factor is 42x.)
Diluent for
CRP
1. Check the quantity of the consumables
(tips and mixing cups).
If consumables are insufficient, supplement or
replace them.
• Loading tips
• Loading mixing cups
NOTE
When performing measurements that
require dilution, at least 2 tips are used.
NOTE
When performing CRP tests that require
dilution at 2x or more, 2 wells (holes) of
mixing cups are used.
2. Load the diluent for CRP.
• Loading diluent
CRP
一般比色
図-4.9
CRP
CM slide
3.
Load a sample and slides.
For details of the sample and slide loading
method, refer to Section 3.3 (3).
NOTE
When measuring a CRP slide and CM
slide at the same time, stack the CRP
slide last or first in the cartridge.
If a CM
slide is measured before or after a CRP
slide, extra tips are used.
3