10-7
10.2 Upper Paper Magazine Table
10
Reinstallation
Reinstallation is essentially in the reverse order of
removal.
NOTE: If the magazine table slider does not run smoothly
after reinstallation, loosen the four screws securing
the processor side bracket and retighten them after
sliding the table in and out several times.
Removal
1. Remove the upper magazine table unit
.
2. Remove the four screws, then the slide table, two
brackets and two magazine guides.
Reinstallation
Reinstallation is essentially in the reverse order of
removal.
Removal
1. Remove:
• Upper magazine setting lever
• Upper magazine setting lever lock arm
• Upper slide table/brackets
2. Remove the three E-rings, then the two bearings,
four collars and camshaft.
LII400
Processor Side Bracket Screws (4)
10.2.5
Upper Slide Table/Bracket Removal/Reinstallation
LII377
Screws (4)
Slide Table
Magazine Guides (2)
Brackets (2)
10.2.6
Magazine Moving Cam Roller Replacement
LII378
Camshaft
Bearings (2)
Collars (4)
E-rings (3)