CR-IR346RU
Service Manual
IN - 26
010-051-00
10.10.2000 FM2887 (1)
IN - 26
7.2.1 Direct Connection of the CL/RU
◆
NOTE
◆
For the I/F cable used for direct connection, a cross cable should be used.
(1) Attach a ferrite bead to the I/F cable.
To do so, wrap the cable once completely around the ferrite bead and tie the cable and
its loop together.
FR6H7725.EPS
Ferrite bead
I/F cable
connector
I/F cable
#1
[Attach] Ferrite bead
#2
[Secure] Cable tie
I/F cable
50–80mm
FR6H7810.EPS
(2) Connect the I/F cable to the I/F cable connector of the RU.
REFERENCE
The I/F cable should be connected with its ferrite bead end on the RU side.
010-051-01