Output signal hold:
Output signal is held during calibration,
processing recoverable sensor, process-
ing diagnosis of sensor, warm-up, PID
auto tuning, under set up maintenance
mode "available" and blowdown. The
hold function can also be released.
Valve and Flow meter (option):
Selects zero or span gas during manual
zero or span calibration. Mounted on the
side of the converter.
Communication function:
RS485 (MODBUS) (option)
Combustion efficiency display (option):
When you select this display, "rich mode
display" will be simultaneously displayed.
This function calculates and displays
combustion efficiency from oxygen
concentration and measured gas tem-
perature.
Thermocouple (R) is required for tem-
perature measurement.
Operating temperature:
−
20 to +55ºC
Operating humidity:
95% RH or less, non condensing
Storage temperature:
−
30 to +70ºC
Storage humidity:
95% RH or less, non condensing
Construction:
Dust-proof, rainproof construction
(corresponding to IP66 or IP67 of IEC)
Material:
Aluminum
case
Outer dimensions (H x W x D):
170 X 159 X 70mm (IP66, Bench type)
220 X 230 X 95mm (IP67)
182 X 163.5 X 70.6mm (Bench type)
Mass {weight}:
IP66: Approx. 2kg (excluding cable and
detector)
IP67: Approx. 4.5kg (excluding cable and
detector)
Finish color:
IP66:
Case:
Silver
Cover: Pantone Cool Gray 1C-F
IP67: Munsell 6PB 3.5/10.5 (blue)
Cover: Silver (case)
Mounting method:
Mounted flush on panel or on pipe
Converter specification (ZKM)
Concentration value indication:
Digital indication in 4 digits
Contact output signal:
(1) Contact specification; 6 points, 1a 250V AC/3A or 30V DC/3A
(2) Contact function;
s 5NDERMAINTENANCE
s 5NDERBLOWDOWN.OTE
s 3PANCALIBRATIONGASVALVE
s :EROCALIBRATIONGASVALVE
s )NSTRUMENTANOMALIES.OTE
s !LARM.OTE
Note1) The following Instrument errors (1) Thermocou-
ples break (2) Sensor break (3) Temperature fault
(4) Calibration fault (5) Zero/span adjustment fault
(6) Output error turn the contact-ON
Note2) Alarm selects just one as mentioned below (1)
High (2) Low (3) Upper and Lower (4) High-high
(5) Low-low, it turns ON while operating.
Note3) Under blow down is available in case of option,
and it turns ON while operating.
Contact input signal:
(1) Contact specification; 3points (the following option)
ON; 0V (10mA or less), OFF; 5V
(2) Contact function;
s %XTERNALHOLD
s #ALCULATIONRESET
s (EATER/&&
s "LOWDOWNOPTION
s )NHIBITIONOFCALIBRATION
s #ALIBRATIONSTART
s 2ANGECHANGE
Calibration method:
(a)
Manual calibration with key operation
(b) Auto. calibration (option)
Calibration cycle; 00 day 00 hour to
99 days 23 hours
(c)
All
calibration
Calibration gas:
s !VAILABLERANGESETTINGS
Zero gas; 0.010 to 25.00% O
2
Span gas: 0.010 to 50.00% O
2
s 2ECOMMENDEDCALIBRATIONGASCONCEN
tration
Zero gas; 0.25 to 2.0% O
2
Span gas; 20.6 to 21.0% O
2
(oxygen concentration in the air)
Blowdown:
A function for blowing out with com-
pressed air dust that has deposited in
the flow guide tube. Blowdown can be
performed for a predetermined time and
at predetermined intervals.
Blowdown cycle; 00 hour 00 minute to
99 hours 59 minutes
Blowdown time; 0 minute 00 second
to 0 minutes 999
seconds
(option)