![Fuji Electric FSC-1 Instruction Manual Download Page 185](http://html1.mh-extra.com/html/fuji-electric/fsc-1/fsc-1_instruction-manual_2341884185.webp)
176
INF-TN3FSC-E
FLOW VELOCITY PROFILE DISPLAY FUNCTION
(OPTION)
Pulse Doppler method enables the analysis and display of
the flow velocity profile in real time. The results can be used
to decide the appropriate measurement location, for flow
diagnosis, and laboratory test.
SPECIFICATIONS
Measuring fluid:
Uniform liquid in which ultrasonic
waves can propagate.
Turbidity of fluid:
Axisymmetric flow in a filled pipe.
Fluid temperature:
−40 to +100°C (FSDP2)
−40 to +80°C (FSDP1,FSDP0)
Air bubble quantity:
0.02 to 15vol% (Velocity is 1m/s)
Pipe size:
Small type sensor : ø40 to ø200mm
Middle type sensor :ø100 to ø400mm
Large type sensor :ø200 to ø1000mm
Measurement range:
0 to ±0.3 … ±Maximum Velocity (de-
pending on the pipe diameter)
Refer to chart, page 145.
Note) This function is to observe flow
velocity profile, and it may be different
from actual flow rate.
DETECTOR FOR FLOW VELOCITY PROFILE
MEASUREMENT (TYPE: FSDP)
Mounting method:
on outside of existing pipe
Ambient temperature:
−20 to +80°C
Ambient humidity:
100% RH or less
Enclosure:
IP67 (waterproof BNC connector
required.)
Material:
Sensor
housing:
PBT
Guide frame:
Aluminum alloy
Mounting belt: Plastic cloth or stain-
less
EU Directive Compliance
LVD (2014/35/EU)
EN
61010-1
EMC (2014/30/EU)
EN 61326-1 (Table 2)
EN 55011 (Group 1 Class A)
EN 61000-3-2 (Class A)
EN
61000-3-3
EN
61326-2-3
RoHS (2011/65/EU)
EN
50581