Fudakang 481109 Instructions For Use Manual Download Page 5

 

5

 

Note:   
1.Any solutions, dilution and cleaning method other than mentioned above might cause   
  reliability problems with the Infrared thermometer. 
2.Contraindications for the following: placing the thermometer probe on scarred tissue   
  or tissue compromised by skin disorders; patients in trauma; 

Patients treated with certain skin’s inflammatory drugs; and placing the thermometer 

probe on skin exposed to direct sunlight, fireplace heat, cold compress therapies, air 

conditioner flow, etc.

   

3. Do not service the unit while the device is in use. 
4.The patient is an intended operator. The patient can measure, transmit data under normal 

circumstances and maintain the device and its accessories according to the user manual.   

5. Please remove the batteries if it is not in use for a long time (More than 3 months).   
6. Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT 

7. Before every use, check the device. Do not use the device if it is damaged in any way. 

The continuous use of a damaged unit may cause injury, improper results, or serious 
danger. 

8. No modification of this device is allowed. 
 

8. Precaution for Measuring Temperature 

Abnormality 

Meaning 

 

Measured result is over 105°F   
(Body measuring mode) 

 

Measured result is below 93.0°F 
(Body measuring mode) 

 

The ambient temperature is beyond the range for proper 
working temperature of 40°F – 104°F 

 
9. Specifications 

 

Model No.

 

FT-F41

 

Measuring Range

 

Forehead 34.0

-43.0

( 93.2

-109.4

)   

Surface 0.0

-100.0

( 32.0

-212.0

)   

 

Measuring time 

minimum  measuring time is 2 seconds; and minimum time between 
measurements for the same site is 2 seconds 

Accuracy 

±0.2

/0.4

   34.0

-34.9

 (93.2

- 94.82

±0.2

/0.4

   35.0

-42.0

 (95.0

-107.6

±0.2

/0.4

   42.1

-43.0

 (107.78

-109.4

 ) 

Resolution

 

0.1

/0.2

 

Operation Condition

 

10.0

-40.0

(50.0

- 104.0

Rh≤80%

 

Pressure: 86~106 kPa

 

Storage Condition

 

-25.0

-55.0

(-13.0

- 131.0

Rh≤95%

 

Pressure: 86~106 kPa

 

Battery type and battery life 

AA 2pcs 
The typical service life of new and unused batteries is 300 measurements   

Protection against electric 
shock 

Internally powered ME equipment 

Summary of Contents for 481109

Page 1: ...RIPTION 1 Table of Contents 2 Product Overview 3 Outline Drawing 4 The Definition of Symbols 5 Measuring 6 Battery Voltage Display Replacement 7 Operation Precautions 8 Precaution for Measuring Temper...

Page 2: ...ading data with the following advantages 1 Easy to clean sensor 2 Fast accurate 3 Convenient 4 32 memories 5 Last memorized temperature display when you turn on the power 3 Outline Drawing 4 The Defin...

Page 3: ...ces 03 through 06 are not active Back light laser selection Press light bulb sunshine button to toggle between the following functions in this order back light on laser on back light off laser off Whe...

Page 4: ...eep the IR Thermometer dry and never immerse it in water or any liquid 4 Do not keep the IR Thermometer in a high temperature environment for a long time and keep away from sunlight dust and dirt 5 Ke...

Page 5: ...device Do not use the device if it is damaged in any way The continuous use of a damaged unit may cause injury improper results or serious danger 8 No modification of this device is allowed 8 Precauti...

Page 6: ...s of oral and rectal methods 2 How to take forehead temperature properly Gently touch the probe to center of forehead or no contact about 5 10 cm distance the probe to center of forehead Make sure rem...

Page 7: ...y this machine should be observed to verify normal operation in the configuration in which it will be used This device complies with Part 15 of the FCC Rules Operation is subject to the following two...