Note:
1.Any solutions, dilution and cleaning method other than mentioned above might cause
reliability problems with the Infrared thermometer.
2.Contraindications for the following: placing the thermometer probe on scarred tissue
or tissue compromised by skin disorders; patients in trauma;
Patients treated with certain skin’s inflammatory drugs; and placing the thermometer
probe on skin exposed to direct sunlight, fireplace heat, cold compress therapies, air
conditioner flow, etc.
3. Do not service the unit while the device is in use.
4.The patient is an intended operator. The patient can measure, transmit data under normal
circumstances and maintain the device and its accessories according to the user manual.
5. Please remove the batteries if it is not in use for a long time (More than 3 months).
6. Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT
7. Before every use, check the device. Do not use the device if it is damaged in any way.
The continuous use of a damaged unit may cause injury, improper results, or serious
danger.
8. No modification of this device is allowed.
8. Precaution for Measuring Temperature
Abnormality
Meaning
Measured result is over 105°F
(Body measuring mode)
Measured result is below 93.0°F
(Body measuring mode)
The ambient temperature is beyond the range for proper
working temperature of 40°F – 104°F
9. Specifications
Model No.
FT-F41
Measuring Range
Forehead 34.0
℃
-43.0
℃
( 93.2
℉
-109.4
℉
)
Surface 0.0
℃
-100.0
℃
( 32.0
℉
-212.0
℉
)
Measuring time
minimum measuring time is 2 seconds; and minimum time between
measurements for the same site is 2 seconds
Accuracy
±0.2
℃
/0.4
℉
34.0
℃
-34.9
℃
(93.2
℉
- 94.82
℉
)
±0.2
℃
/0.4
℉
35.0
℃
-42.0
℃
(95.0
℉
-107.6
℉
)
±0.2
℃
/0.4
℉
42.1
℃
-43.0
℃
(107.78
℉
-109.4
℉
)
Resolution
0.1
℃
/0.2
℉
Operation Condition
10.0
℃
-40.0
℃
(50.0
℉
- 104.0
℉
)
Rh≤80%
Pressure: 86~106 kPa
Storage Condition
-25.0
℃
-55.0
℃
(-13.0
℉
- 131.0
℉
)
Rh≤95%
Pressure: 86~106 kPa
Battery type and battery life
AA 2pcs
The typical service life of new and unused batteries is 300 measurements
Protection against electric
shock
Internally powered ME equipment