Sample Extractor Assemblies
FTS Systems LyoStar™ 3
28
Rev 006, 01/14
© SP Scientific 2014
Sample Extractor Installation
Standard Door Removal
Before installing the Sample Extractor Assembly, the standard acrylic door must be
removed from the freeze-dryer.
Ensure that the unit is not in operation and that the product chamber is not
1.
under vacuum.
Open the product chamber door. Have an assistant in position to support the
2.
door.
Remove the upper and lower quick release pins from the hinge assembly, and
3.
remove the standard door.
Sample Extractor Installation
Prepare the Sample Extractor Assembly door for installation by unscrewing
1.
and removing both thumbscrews from the front panel.
Remove the front antechamber plate and selector arm.
2.
Quick Release Pins
Thumbscrew
Thumbscrew
Summary of Contents for LYOSTAR 3
Page 2: ......
Page 6: ...iv Rev 006 01 14 SP Scientific 2014...
Page 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Page 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Page 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Page 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Page 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Page 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Page 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Page 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Page 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Page 81: ......