Section 3
Installation
14
FT702LT/D Flat Front Wind Sensor Manual
3 INSTALLATION
See safety instruction requirements on page 4.
The wind sensor installation must be properly designed to ensure the correct
operation of the sensor. This section is for guidance only. It is the responsibility of the
designer and installer to ensure that the installation and its design is fit for purpose.
Please see disclaimer section 1.5.
3.1 Connectivity
3.1.1
Connector Details
All electrical connections are made to the FT702LT/D via an 8-way multipole connector located in the base of
the wind sensor housing. The wind sensor connector pin designations are shown in Figure 10 and the mating
connector manufacturer’s part numbers in Figure 11.
Figure 10: Sensor connector pin out
Connector Type
Connector
Description
Manufacturers Part Number
FT Sensor
Type
Maximum Outer
Cable Diameter
Manufacturer
FT702LT/D
Cable Side
Connector
8 way plug
SS104Z129-1
FT702LT/D
Analogue
Option
8.0mm
W.W.Fischer
FT702LT/D
Cable Side
Connector
8 way plug
SX2F1C-P08NJH9-0001
FT702LT/D
Analogue
Option
9.2mm
ODU
Figure 11: Cable Connector Sourcing Options
3.1.2
Cable Details
The mating connectors for the FT702LT/D are suitable for use with cables with overall diameters of up to 9mm
and for individual cores of diameters of up to 1.2mm. Cable such as SD980CPTP
3x2x0.5mm2 from SAB
Brockskes or similar types may be used. Care must be taken to ensure that the cable is suitable for the
environment it will be used in and is adequately approved, for example AWM Style 21198.
In an area with a moderate or severe lightning strike exposure the cable shield will not provide sufficient
protection. In this case the cable will require further shielding such as being enclosed in a metal pipe or conduit
(see section
3.3).