Isolation Systems Inc
Operation and Maintenance Manual PASS-1600ISO
2 October, 2020
Isolation Systems Inc
11
889 Erie Ave, North Tonwanda NY, 14120
2.5 Example Configurations
The PASS-1600ISO Negative Pressure
Shelter Example
The unit is placed outside the Isolation
shelter, the large air "intake" duct is
connected to the shelter duct port. A
small intake duct is supplied to the anti-
room and provides a slight negative
pressure to the ante-room.
The large exhaust duct is routed from
the PASS exhaust and is vented to the
outside to create negative pressure for
the room.
NOTE: Ducts not shown are not connected
Large Intake Duct
Large Exhaust Duct
Small Intake Duct
Small Exhaust
PASS1600ISO
Controller/Pressure Tube
Air Flow
INTAKE
EXHAUST
Ante-room
Intake
Controller
Exhaust
Unit Inside Shelter
Negative Pressure with Ante-room
Set-up
Large Intake: Patient Room, No Duct Small
Intake: Ducted to Ante-room
Large Exhaust: Patient Room, No Duct
Small Exhaust: Ducted outside
Dampers: Exhaust 1/3 open, Intake closed
Controller: Pressure tube outside shelter
Room