37
- Press the
HISTORY
key for 8 second while in normal mode to enter
the Calibration Mode, and the wind factor digits will start flashing.
You can skip over any setting by press the
SET
key. Press
+
key or
MIN/MAX
key to select the units or scrolls the value. Holding the
+
key or
MIN/MAX
key for 2 second will increase/decrease digits in
great steps. To exit the Calibration mode at any time, press the
HISTORY
key.
Wind Speed Calibration (default is 1.0, adjustment rang 0.75 to
1.25)
Rainfall data Calibration (default is 1.0, adjustment rang 0.75 to
1.25)
History rainfall Calibration
Outdoor humidity Calibration
Outdoor temperature Calibration
Indoor humidity Calibration
Indoor temperature Calibration
Wind Speed Calibration
Wind speed is the most sensitive to installation constraints. The
weather station should not be located close to buildings, trees or other
obstructions.
Many installations are not perfect and installing the weather station on
a roof can be difficult. Thus, you can calibrate for this error with a wind
speed multiplier.
In addition to the installation challenges, wind cup bearings (moving
parts) wear over time.
Without a calibrated source, wind speed can be difficult to measure.
We recommend using a calibrated wind meter and a constant speed,
high speed fan.
Temperature Calibration
Temperature errors can occur when a sensor is placed too close to a
heat source (such as a building structure, the ground and when placed
in direct sunlight without proper shielding in hot weather environments).