6.1.6 Closing the cutting mill
1.
Before closing the cutting mill, clean the cutting chamber, the con-
tact surfaces of the housing and, in particular, the locking surfaces
of the lock.
2.
Swivel the upper part of the housing (2) slowly closed until it is
resting on the lower part of the housing (10).
3.
Close the closing lid (6).
4.
Lock the latch clamp (13).
5.
Switch the control switch (19) on the back of the device to HAND.
6.1.7 Checking if the rotor is turning freely
n
To check if the rotor is turning freely, the closing lid must be closed.
‘Setting the gap width of the knives’ on page 31).
n
Remove the cover cap (Z) of the central bore hole in the closing lid.
n
Guide the hexagon socket screw key provided through the central
bore hole in the closing lid.
n
Now turn the hexagon socket screw key to check if the rotor is
turning freely.
Using the device
- 34 -