9.
Tighten the 3 cylinder-head screws for adjusting the grinding gap
using the hex wrench (c).
10.
Remove the feeler gauge strip (d) from the grinding chamber.
11.
Refit the device's belt guard (6).
12.
Then close the lid (7).
13.
Push in the source material container (8).
6.3.4 Attaching the support walls
It is practically only necessary to replace the support
walls (11) when the crushing plates (17, 21) are
replaced, because their material needs to be changed
and even minimal amounts of abrasion on the support
walls are to be avoided.
Tool
c
Hex wrench - size 6
f
Socket wrench - size 13
g
Hex wrench - size 4
h Open-ended wrench - size 13
i
Hex wrench - size 5
n Cross-head screwdriver
Using the device