4.4 Ambient conditions
WARNING!
Mains voltage!
– The device may only be operated indoors.
– The surrounding air may not carry any electrically
conductive dust.
– Maximum relative humidity 80% for temperatures
up to 31°C, linearly decreasing down to 50% rela-
tive humidity at 40°C.
n
The room temperature has to stay between 5 - 40°C.
n
Altitudes up to 2000 m
n
Degree of pollution 2 according to IEC 664.
4.5 Electrical connection
Before establishing the connection, compare the voltage and current
values stated on the type plate with the values of the mains system to be
used.
CAUTION!
Ignoring the values on the type plate may result in
damage to the electrical and mechanical components.
4.5.1 Adjusting the mains voltage
4.5.1.1
Version SPARTAN and PRO (valid till serial number 05727)
CAUTION!
If the values 06 or 07 are set at a voltage of 230V~, it
will result in a defect in the sieve shaker during opera-
tion. Failure to observe this will render void the guar-
antee, and releases us from liability for any resulting
damage to the device.
NOTICE!
Only qualified personnel may change the voltage range
on the device!
Installation