d MANUFACTURER’S declaration – electromagnetic IMMUNITY 

Guidance and manufa

cturer’s declaration – electromagnetic immunity

 

The 

T688B Series Chair

 is intended for use in the electromagnetic environment specified below. The 

customer or the user of the 

T688B Series Chair

 should assure that it is used in such an environment. 

IMMUNITY test

 

IEC 60601 TEST 
LEVEL

 

Compliance 
level

 

Electromagnetic environment 

– 

guidance

 

 
 
 
 
 
 
 
Conducted RF 
IEC 61000-4-6 
 
 
 
 
 
 
Radiated RF 
IEC 61000-4-3 

 
 
 
 
 
 
 
3 Vrms 
150 kHz to 80 
MHz  
 
 
 
 
 
 
3 V/m 
80 MHz to 2,5 
GHz 

 
 
 
 
 
 
 
 
3Vms 
 
 
 
 
 
 
 
3 V/m 
 

Portable and mobile RF communications 
equipment should be used no closer to 
any part of the 

T688B Series Chair

including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of 
the transmitter.

  

 
Recommended separation distance 
 

 

 
where 

is the maximum output power 

rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and 

is the recommended 

separation distance in metres (m). 
 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey,

 a

 should be 

less than the compliance level in each 
frequency range.

 b 

 
Interference may occur in the vicinity of 
equipment 
marked with the following symbol: 
 

 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption 
and reflection from structures, objects and people. 

a

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 

and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be 
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the 
location in which the [ME EQUIPMENT or ME SYSTEM] is used exceeds the applicable RF compliance 
level above, the [ME EQUIPMENT or ME SYSTEM] should be observed to verify normal operation. If 
abnormal performance is observed, additional measures may be necessary, such as re-orienting or 
relocating the [ME EQUIPMENT or ME SYSTEM]. 

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [

V

1] V/m. 

 
 

Summary of Contents for T688B Series

Page 1: ...Medical Treatment Chair T688B Series Instruction Manual ...

Page 2: ...ND CONTROLS PROGRAMMING 7 3 3 1 Hand Control 7 3 3 2 Programming Instructions 8 3 3 3 Recalibration 8 3 MAINTENANCE 9 4 1 INSPECTION SERVICING 9 4 2 CLEANING 9 4 3 FABRIC STANDARDS 10 4 4 TROUBLE SHOOTING 10 4 ARRANGING A SERVICE 11 5 WARRANTY INFORMATION 12 6 WASTE DISPOSAL 12 7 PHYSICAL DESCRIPTION 13 8 1 DESCRIPTION OF DEVICE 13 8 2 TECHNICAL SPECIFICATIONS 13 8 2 1 Dimensions Weight Capacity 1...

Page 3: ...the seat and leg rest The chair is fitted with castors which are specifically used to aid cleaning and or positioning of the chair within the room The T688B Series chair shall not be used to transport patients or any other items Ensure that the backrest is fully elevated before moving the chair For models fitted with dual hand controls the programmable unit Nurse s Hand Control is for the use of t...

Page 4: ...ntraindication for medical treatment chairs But limitation of use in patients with history of Allergy to covering material of chairs Pressure sores 1 5 Reporting serious incident Any serious incident that has occurred in relation to this medical device shall be reported to the manufacturer and local health authority in accordance with the local regulations Within the EU the user must report any se...

Page 5: ...as per instruction 5 above 9 Arms maybe swung out or lowered flat depending on style to seat level to facilitate side transfer of patient Raise or lower the seat height to prevent unnecessary bending or lifting by staff members 10 IV Pole if fitted can be removed or repositioned by loosening the knob at the lower end of the mounting bracket and retightening the knob 11 Lift up tray if fitted can b...

Page 6: ...mergency stop press any button on either the nurse hand control or the patient hand control if fitted 20 USB charger if fitted can be used to charge a USB device such as a mobile phone a music player or a camera Maximum output of the charger is 4 watt NOTE It may not charge all USB devices depending on the power input requirement of the USB devices NOTE To use the USB charger ensure the chair is p...

Page 7: ...witched on Refer to section 3 2 The backup batteries if fitted will recharge automatically and slow to trickle charge when full It is important to charge the backup battery for a minimum of 24 hours before use the chair 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZARD CAUSE PREVENTA...

Page 8: ...please contact Fresenius Medical Care or your local representative at the details provided in section 5 The battery is not user replaceable The chair must be connected to the power supply to recharge at least every 5 days 3 3 HAND CONTROLS PROGRAMMING Each T688B Series Medical Treatment Chair can be supplied with two hand controls One nurse operated hand control and one patient operated hand contr...

Page 9: ...calibrateT688B Series chair as follows using the nurse s hand control 1 Press both up and down arrows of the Height function simultaneously and hold until the beeping stops approximately five seconds NOTE both UP and Down buttons must be activated at exactly the same time 2 Press both up and down arrows of the Backrest control function simultaneously and hold until all the actuators have stopped m...

Page 10: ...material or a sample provided by the manufacturer The following should be used as a guideline only and no claims will be accepted for damage as a result of following these guidelines RECOMMENDED CLEANING PROCEDURE FOR MEDICAL TREATMENT CHAIRS The following recommended procedure should only be used if it generally conforms to Infection Control guidelines of the hospital Step 1 Bodily fluids may car...

Page 11: ...n correctly as indicated by the green indicator light on the nurse s hand control 3 Ensure that the batteries have been on charge for 24 hours minimum This should occur at least every 5 days 4 Ensure there are no nicks or cuts in the hand control or actuator cables 5 Determine what is not operating as follows The leg motor vertically mounted under the seat Depress the hand control button for the l...

Page 12: ...ote Please quote the chair serial number which can be found on the Type Label attached to the chair An example of the Type Label is in section 8 2 7 CONTACT INFORMATION https fmc seating com MANUFACTURER SERVICE CENTRE FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road Scoresby VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL 61 0 3 9780 9509 SALES TEL 61 0 3 9780 9500 ...

Page 13: ...onth warranty 6 The T688B Series Medical Treatment Chairs are not covered by warranty if used outdoors 7 The use of unauthorised labour will void the warranty Contact Fresenius Medical Care for approval prior to commencing work 8 If you are unsure about any aspect of this warranty please contact Fresenius Medical Care at the number given in Section 5 of this manual If you need to arrange a service...

Page 14: ...g castors or central locking castors removable upholstery of adjustable head rest backrest seat cushions and leg rest metal frame 8 2 TECHNICAL SPECIFICATIONS 8 2 1 Dimensions Weight Capacity T688B Dimensions depending on options Height 128 135 cm Width 112 cm Length 100 cm Weight depending on options 128kg 282 lbs Capacity 300kg 661 lbs 8 2 2 Electrical Safety CB 6 Control Box Specifications Degr...

Page 15: ...V2 Apr 2021 8 2 4 Operating Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 5Transport Storage Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 6 Electromagnetic or other interference Switch off devices which can cause electromagnetic interference ...

Page 16: ...rticle Number Catalog Number Date of manufacture Serial Number CE mark Manufacturer For indoor use only Caution warning Consult Instruction Manual authorized representative in the European Community Medical Device Degree of protection against electrical shock Type B Storage and operating temperature 5ºC 40ºC Storage and operating relative humidity 20 90 ...

Page 17: ...tery is in normal use and requires charging LED Amber ON and LED Green ON indicates mains is on and battery is charging LED Amber Off and LED Green Off indicates Mains is not on and battery requires charging Chair will not operate LED Amber blinking and LED Green blinking and indicates a system error Note When the chair is not connected to the mains press any button on the nurse hand control to wa...

Page 18: ...ns test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The T688B Series Chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any Interference In nearby electronic equipment RF emissions CISPR 11 Class B The T688B Series Chair is suitable for use in all RF emissions establishments other than domestic a...

Page 19: ...wer quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT...

Page 20: ...e in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propa...

Page 21: ... to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 3 691 1 17 1 17 2 33 1011 67 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recomme...