16
Power Failure
What happens when power is restored?
• Your cycler will go through a “Power Fail Recovery” process if power fail happened during
drain, fill, dwell or pause
• An audible beep will be heard. Press
STOp
to mute the warning.
• Your cycler will perform a safety check. The status bar will state “Power Fail Recovery”
and the screen will read “Power Fail Detected Recovery Check in progress”.
• If your cycler passes the safety check it will beep again.
Press
STOp
to mute and then you may select “Resume Treatment”.
• If your cycler does not pass the safety check, select “Cancel Treatment”
and follow the prompts on the screen to start your treatment over.
Use new supplies for new treatment.
NOTE:
If the power has been off for more than
8 hours do NOT resume treatment. Discuss with
your PD Nurse any special instructions in case of
a power failure.