8
P/N 450368-03 Rev. E
D
IAGRAM
1:
F
RONT
P
ANEL
B
UTTONS AND
I
NDICATOR
L
IGHTS
/C
YCLE
P
ROGRAM
O
VERVIEW
Dissolution Cycle
consists of the following Operations:
•
FILL –Input water Fill Valve Opens and allows water into tank. Once water reaches Mid Level Sensor the unit switches
to ADD GRANULES Operation.
During the FILL Operation a Short Rinse Program is done before the water reaches
the Mid-Level Sensor.
•
ADD GRANULES – The indicator light is flashing and the unit is waiting for operator interaction. Once Dry Acid Powder
is added the operator is to press START Button. The unit proceeds to the MIX Operation.
•
MIX – The Mixer Motor turns on for 35 minutes, and then the unit automatically proceeds to DEAERATION Operation.
•
DEAERATION – The Unit sits motionless to allow excess bubble to rise out of solution for 2 minutes. The unit
automatically proceeds to the FINAL FILL Operation.
•
FINAL FILL – The Unit opens the Fill Valve and Closes the Fill Valve once the Final FILL Sensor is contacted by the rising
level in the tank. The program shall continue on to the HOMOGENIZE Operation.
•
HOMOGENIZE – The Mixer Motor runs to mix the contents of the tank for 10 minutes and then the program continues
to the TRANSFER Operation.
•
TRANSFER – The indicator light is flashing. Unit is waiting for Operator Interaction. Once Transfer Operation is
complete the program continues to the CYCLE COMPLETE Operation.
•
CYCLE COMPLETE – The Unit will sit motionless until next Operator Interaction is initiated.
Rinse Cycle
consists of the following Operations:
•
1
st
FILL –Input Water Fill Valve Opens and allows water into tank. Once water reaches 25-gallon sensor the unit
switches to RECIRCULATE Operation.
•
1
st
RECIRCULATE – The RECIRCULATE Operation runs for a 12 minute period. The spray ball and mixer motor runs for
duration of the time. Then, the program continues to DRAIN Operation.
•
1
st
DRAIN– The DRAIN Operation drains the contents of the tank. Then, the program continues to the 2
nd
FILL
Operation
•
2
nd
FILL – Input Water Fill Valve Opens and allows water into tank. Once water reaches 25-gallon sensor the unit
switches to RECIRCULATE Operation.
•
2
nd
RECIRCULATE – The RECIRCULATE Operation runs for 12 minute period. The spray ball and mixer motor runs for
duration of the time. Then, the program continues to DRAIN Operation.
•
2
nd
DRAIN – The DRAIN Operation drains the contents of the tank. Then, the program continues to CYCLE COMPLETE
Operation.
•
CYCLE COMPLETE – The Unit shall sit motionless until Operator Interaction is initiated.
*PAUSE State and STEP Mode: For information on the use of PAUSE State and STEP Mode refer to Section 9: “Manual
Control Operations.”
RINSE Operation status
Indicator Lights
Dissolution Operation status
Indicator Lights
Rinse Cycle START
Push Button
*STEP Push Button
*
STEP Mode ON/OFF
Push Button
*PAUSE Push
Button
Time Remaining
Indicator
Indicator Lights
Dissolution Cycle START
Push Button
Summary of Contents for GranuFlo 450368-03
Page 2: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 4: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 6: ... PAGE INTENTIONALLY LEFT BLANK ...
Page 39: ...P N 450368 03 Rev E 33 PAGE INTENTIONALLY LEFT BLANK ...
Page 41: ...P N 450368 03 Rev E 35 PAGE INTENTIONALLY LEFT BLANK ...
Page 48: ...42 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 49: ...P N 450368 03 Rev E 43 APPENDIX A DRY ACID MIXING INSTRUCTIONS ...
Page 50: ...44 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 51: ...P N 450368 03 Rev E 45 ...
Page 52: ...46 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 53: ...P N 450368 03 Rev E 47 ...
Page 54: ...48 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 55: ...P N 450368 03 Rev E 49 ...
Page 56: ...50 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 57: ...P N 450368 03 Rev E 51 ...
Page 58: ...52 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 59: ...P N 450368 03 Rev E 53 Appendi x C Dry Acid Mixing Instructions ...
Page 60: ...54 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 61: ...P N 450368 03 Rev E 55 ...
Page 62: ...56 P N 450368 03 Rev E PAGE INTENTIONALLY LEFT BLANK ...
Page 63: ...P N 450368 03 Rev E 57 ...
Page 65: ...P N 450368 03 Rev E 59 PAGE INTENTIONALLY LEFT BLANK ...
Page 67: ...P N 450368 03 Rev E 61 PAGE INTENTIONALLY LEFT BLANK ...