2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 517
BL- 4.4.4 ISOLATE BLOOD LEAK DETECTOR
a)
IMPORTANT! To prevent damage turn the
machine off!
b)
Figure right
, at the distribution board, between
and below positions #7 and #9, unplug the Blood
Leak ribbon Cable!
c) Using a flashlight, check the vacant connector. If
‘white’ corrosion or damaged
male
pins are located
the distribution board may need to be replaced
!
d) Leaving the cable unplugged, return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM)!’
e) Call debug screen 4. Is
LEAK
AND
DIMN
BOTH = 0.0?
Yes
LEAK
and
DIMN
BOTH = 0.0! See procedure number
(page 517).
No
LEAK
and/or
DIMN
are more than 0.0! THREE (3) possible bad components (Component List
below). Swap the components in, one at a time, and in between repeat procedure number
(page 517), parts d and e, to test each new component until
LEAK
and
DIMN
are BOTH
0.0!
Component List:
1)
Sensor Board cable;
2)
Sensor Board
1
(see
3)
Distribution board.
1
To prevent a “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
)).
BL- 4.5.0 RETURN BLOOD LEAK SYSTEMS
a)
IMPORTANT! To prevent damage turn the machine off!
b) Return the Blood Leak ribbon cable to the distribution board.
c) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
d) Allow six (6) minutes for stabilization!
e) ENSURING the external flow indicator’s ‘bob’ is rising and falling see procedure number
518).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...