2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 467
TMP- 4.2.0 ISOLATE SECONDARY CIRCUIT
a) Open the shunt door to close valve #24 and #25. This isolates the external dialysate lines (see
b) Attempt to achieve -250 mmHg then clamp the syringe tubing. Can you achieve -250 and HOLD it
(+/- 15 mmHg per minute) now?
Yes
-250 mmHg achieved and HOLDS! Proceed to
page 468
, procedure number
No -250 cannot be achieved and / or held! See procedure number
(page 467).
Figure 71 C - Isolation Area
(excludes dialysate lines)
TMP- 4.3.0 -250 LEAKING / ISOLATE DIALYSATE LINES
a) Trace the to and from dialyzer lines to where they attach to the rear
of the machine.
b) INSIDE the machine, clamp the clear (non-braided) tubing, at BOTH
dialysate line connectors.
c) Can you achieve -250 mmHg and HOLD it (+/- 15 mmHg per minute)
now?
Yes
-250 mmHg achieved and HOLDS!
IMPORTANT!
Remove the clamps then see
(page 468).
No --250 cannot be achieved and / or held! FIVE (5) possibilities,
1)
Leaking dialyzer quick
connectors and/or O-rings;
2)
Leaking external flow indicator fittings;
3
) Leaking optional
fluid sample port;
4)
Leaking dialyzer line connections at the rear of the machine;
5)
Leaking external dialysate line filter O-ring/housing.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...