2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 421
FIL- 2.1.0 NO AIR SEEN BUT FILACT = 1 / ISOLATE SUPPLY VOLTAGES
a) Call debug screen 1, TWO (2) voltage checks:
Check #1: 5V Est. (right column):
Between
4.7 and 5.3?
Check #2: 12V Est. (right column):
Between
11.7 and 12.3?
b) Are BOTH in range AND stable i.e. not changing more than 0.1 per minute?
Yes
BOTH 5V and 12V within range AND stable! See procedure number
(page 421).
No One or both NOT in range OR unstable! See parts A) through C) below:
A) To avoid pulling cables loose, GENTLY open the card cage.
B) Behind the card cage, ENSURE the 24V POWER harness has remained plugged in.
C) Proceed to
page 599
, procedure number
FIL- 2.3.0 VOLTAGES IN RANGE / ISOLATE AIR SENSOR CIRCUIT
a) FIGURE BELOW, inside the distribution board, unplug the female Air Sensor’s connector i.e.
4
th
CONNECTOR CAP,
5
th
position from the left. If present, DO NOT unplug the CBE board!
b) Using a flashlight, ENSURE no corrosion or damaged pins inside the distribution board!
c)
Per the Figure next page
, place a resistor plug, from the
FOUR-RESISTOR SET
into the Air
Sensor’s distribution board position!
If CBE equipped, using a flashlight, ENSURE the resistor
covers ALL CBE board pins, especially the top pin!
Parts d and e next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...