2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 369
CO- 8.3.0 BALANCING CHAMBER VOLUME (BCV)
Figure below, the
B
alancing
C
hamber(s)
V
olume (BCV) MAY HAVE BEEN factory recorded, a three-digit
number (example
61.3
), written on the top or side of the balancing chambers (Yes or No)?
Yes
BCV is recorded! See
procedure number
(page 369).
No BCV is NOT recorded!
Proceed to
page 370
procedure number
CO- 8.3.1 BCV IS RECORDED
Call debug screen 3.
BCV
(upper right) is the stored
B
alancing
C
hamber
V
olume, of
BOTH chambers (
example:
613 = 61.3 ml). Does screen 3’s
BCV
= the factory recorded
value?
Yes
BCV
= the factory volume! See procedure number
(page 370).
No Read parts a THROUGH f below BEFORE performing them!
a) Perform BALANCING CHAMBER VOLUME CALIBRATION, per the CALIBRATION
PROCEDURES booklet,
BUT enter the factory recorded volume
!
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) Call debug screen 3. If the calibration was performed correctly
BCV
= the factory value.
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
e) Allow fifteen (15) minutes BEFORE continuing to part f!
f) Is the external meter’s reading now /- 0.4 mS of TCD?
Yes
Conductivity /- 0.4 mS of TCD.
BCV
was incorrect! Do NOT continue!
(page 370).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...