2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 320
CO- 1.0.1 CONDUCTIVITY IS LOW / ISOLATE BICARB PUMP CONTROL
Call debug screen 0. Figure right, what color is the
Bic
pump’s ‘dot’, BLUE or
WHITE?
1)
IF BLUE:
Proceed to
page 321
, procedure number
2) IF WHITE:
a)
Select the ‘Dialysate’ tab to call the Dialysate screen.
b) Figure below, is the
[Bicarbonate]
window MORE THAN zero (0)?
Yes
[Bicarbonate]
window more than zero (0)!
Proceed to
page 374
No
[Bicarbonate]
window
= zero (0)! Acetate concentrate is selected! See
(page 320).
Figure 56 – Dialysate Screen
CO- 1.0.2 [Bicarbonate] WINDOW = 0 / ACETATE IS SELECTED!
1:34 Acetate is rarely used and uses ONLY one concentrate, ACETATE. Troubleshooting must be
performed using bicarb i.e. the
[Bicarbonate]
window MUST be more than zero (0). See parts a THROUGH
c below:
a) Firmly connect to
known good ACID AND bicarbonate!
b) Select the screen’s ‘Conc’ button THEN select the
ACID
that is attached to the machine.
c) Press ‘CONFIRM’! ENSURING the
[Bicarbonate]
window is now MORE THAN zero (0), call debug
screen 0. What color is the
Bic
pump’s ‘dot’ now, BLUE or WHITE?
1)
IF BLUE:
ENSURING [Dialysate Flow] is on, allow five (5) minutes. If [
Conductivity
] remains
less than 13.0 mS see procedure number
2) IF WHITE:
Proceed to
page 374
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...