
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 196
A resistor plug IS NOT ALREADY in place continued:
a) Unplug the female Air Sensor’s connector,
4
th
connector cap
5
th
position, from the LEFT!
b) Call debug screen 0. Figure right, the text box above Chamber #69
MUST say “Air”!
c) Place one of the plugs, from the
FOUR-RESISTOR SET
,
into the Air Sensor’s distribution board position!
*
If CBE modified the resistor plugs into the CBE board pin for pin. ENSURE the TOP CBE
board pin is covered by the resistor
d) If the resistor plug was placed properly, Chamber #69 MUST
now say “No Air”!
(page 196).
VE- 1.0.0 INITIALIZE ACTUATOR-TEST BOARD VALVE CYCLES
a)
Turn the machine OFF!
b) Turn the machine ON and return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 300 ml/min.
(page 197).
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...