72
7.2
Troubleshooting Guide
Description
Cause
Recommended Action
End of infusion detected too early (at
approximately 10 mL).
No end of infusion pre-alarm and alarm
The installed syringe doesn't
correspond to the selected
syringe.
Replace or confirm the right syringe brand.
Calibrate the displacement sensor using
Agilia Partner maintenance software.
Flow rate or displacement drift.
The installed syringe doesn't
correspond with the selected
syringe.
The position sensor calibration
values have drifted.
Replace or confirm the right syringe brand.
Check the displacement calibration.
If necessary, recalibrate the displacement
sensor using Agilia Partner maintenance
software.
Replace the displacement sensor.
See section 8.12, page 115.
Occlusion alarm after the pump has been
turned on.
Wrong calibration of the force
sensor.
Force sensor is out of order.
Flexible circuit is cut.
Recalibrate the force sensor using
Agilia Partner maintenance software.
Replace the force sensor and ribbon cable kit.
See section 8.11, page 110.
Occlusion alarm during the infusion.
The pressure limit selected is too
low.
Wrong calibration of the force
sensor.
Flexible circuit is cut.
Select a higher pressure limit (Pressure
management menu
).
Recalibrate the force sensor using
Agilia Partner maintenance software.
Replace the force sensor and ribbon cable kit.
See section 8.11, page 110.
Disengagement alarm after the pump has
been turned on or during the infusion.
The disengagement micro-switch
is defective.
Carriage flexible circuit is cut.
Replace the disengagement flexible circuit.
See section 8.10, page 105.
Unjustified alarm involving plunger head
position.
Optical switch and/or anti-siphon
arm finger detection are defective.
Flexible circuit is cut.
Check the anti-siphon system.
Check the plunger driver internal mechanism.
Replace the force sensor.
See section 8.11, page 110.
Unjustified alarm involving syringe barrel
clasp and/or syringe body.
Defective syringe barrel clasp
potentiometer.
Flat ribbon cable is defective.
Switch or flat ribbon cable flange
detection is defective.
Check the syringe barrel clasp.
Run test 7. See section 6.2.8, page 47.
If the error remains, replace the syringe barrel
clasp. See section 8.7, page 94.
Calibrate the syringe barrel clasp using
Agilia Partner maintenance software and
check the pump functionality.
Display defect: LED, LCD display.
Control transistors, LED and/or
connections are defective.
Check the display board.
Check the connectIons between CPU board
and display board.
Replace the display board/LCD display.
See section 8.3, page 83.
The power supply indicator doesn't light
up.
Battery is totally discharged.
Power supply board is defective.
Connect the pump to the AC power supply.
Replace the battery.
See section 8.1, page 79.
Check the power cord and AC power supply.
Replace the power supply board.
See section 8.5, page 88.
Battery alarm even though it has been
completely charged.
The power supply board is
damaged.
Replace the power supply board.
See section 8.5, page 88.
The pump turns ON or OFF by itself.
Defective keypad.
Power supply board is damaged.
Replace the upper case. See section 8.2,
page 81.
Check the keypad by running test 10.
See section 6.2.11, page 51.
Replace the power supply board if necessary.
See section 8.5, page 88.
Some keys of the keypad do not work.
Defective keypad.
Replace the upper case.
See section 8.2, page 81.
Check the keypad by running test 10.
See section 6.2.11, page 51.
Summary of Contents for Agilia SP
Page 139: ...139 17 Spare Parts Catalog Refer to the spare parts catalog of the relevant pump ...
Page 143: ...143 VTBI Volume to Be Infused W Watts Term Description ...
Page 147: ...Local Contacts for Servicing Write your contacts here ...
Page 148: ...11378 2_tm_agilia_sp_eng ...
Page 305: ...Local Contacts for Servicing Write your contacts here ...