95
Note: 1 bar = 750 mmHg = 100 kPa = 14.5 PSI.
14.9
Accuracy
14.9.1 Flow Rate Accuracy
14.9.2 Effects of Pressure Variations on Accuracy
Format
Range of Settings
Default
Value
Minimum
Increment
3
L
ev
els
Low
mmHg
50
300 100
50
Medium
mmHg
150
700
250
50
High
mmHg
250
900
500
50
Va
ri
ab
le
Full Range
mmHg
50
900
500
25
50
(50
250)
(250
900)
Maximum
Limit
mmHg
500
900
900
50
DPS
Raise
Threshold
mmHg
50
400
100
50
Drop
Threshold
mmHg
100
400
100
50
Warning
Accuracy (flow rate, time, volume infused, pressure) can be influenced
by syringe model, syringe configuration, extension set configuration,
fluid viscosity, and fluid temperature.
Note: All tests below are in accordance with the IEC 60601-2-24 standard and
ANSI/AAMI ID26.
Values are representative of syringes used during internal tests and are provided
as indicators only.
Accuracy
Flow Rate
± 3%
Back Pressure
Accuracy
(From Mean Values)
+ 39.9 kPa
+ 13.33 kPa
- 13.33 kPa
~ - 3%
~ - 1.5%
~ + 1.5%