995066 - Revision A
Page 24 of 41
9.6 – Leg Actuator
Within this section it will explain the correct procedure on removing and reinstalling the Leg
Actuator for servicing procedures or replacement.
Removal
Step 1 – Remove the Leg Actuator Cable. (Refer to section 9.2.3)
Step 2 – Using a 6mm Allen Key, remove the two bolts securing the Actuator to the Frame.
Step 3 – Actuator can be removed from the Stand Aid.
Refitting / Replacement
Step 4 - Refitting is a reversal of the removal process noting the following point:
A)
Ensure the Actuator is fitted in the correct orientation, with the label facing the floor
as shown in the image above.
B)
Ensure Loctite 243 is applied to the bolts that secure the Actuator.